Design, calibration and validation of a novel 3D printed instrumented spatial linkage that measures changes in the rotational axes of the tibiofemoral joint.

نویسندگان

  • Daniel P Bonny
  • M L Hull
  • S M Howell
چکیده

An accurate axis-finding technique is required to measure any changes from normal caused by total knee arthroplasty in the flexion-extension (F-E) and longitudinal rotation (LR) axes of the tibiofemoral joint. In a previous paper, we computationally determined how best to design and use an instrumented spatial linkage (ISL) to locate the F-E and LR axes such that rotational and translational errors were minimized. However, the ISL was not built and consequently was not calibrated; thus the errors in locating these axes were not quantified on an actual ISL. Moreover, previous methods to calibrate an ISL used calibration devices with accuracies that were either undocumented or insufficient for the device to serve as a gold-standard. Accordingly, the objectives were to (1) construct an ISL using the previously established guidelines,(2) calibrate the ISL using an improved method, and (3) quantify the error in measuring changes in the F-E and LR axes. A 3D printed ISL was constructed and calibrated using a coordinate measuring machine, which served as a gold standard. Validation was performed using a fixture that represented the tibiofemoral joint with an adjustable F-E axis and the errors in measuring changes to the positions and orientations of the F-E and LR axes were quantified. The resulting root mean squared errors (RMSEs) of the calibration residuals using the new calibration method were 0.24, 0.33, and 0.15 mm for the anterior-posterior, medial-lateral, and proximal-distal positions, respectively, and 0.11, 0.10, and 0.09 deg for varus-valgus, flexion-extension, and internal-external orientations, respectively. All RMSEs were below 0.29% of the respective full-scale range. When measuring changes to the F-E or LR axes, each orientation error was below 0.5 deg; when measuring changes in the F-E axis, each position error was below 1.0 mm. The largest position RMSE was when measuring a medial-lateral change in the LR axis (1.2 mm). Despite the large size of the ISL, these calibration residuals were better than those for previously published ISLs, particularly when measuring orientations, indicating that using a more accurate gold standard was beneficial in limiting the calibration residuals. The validation method demonstrated that this ISL is capable of accurately measuring clinically important changes (i.e. 1 mm and 1 deg) in the F-E and LR axes.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Optimized design of an instrumented spatial linkage that minimizes errors in locating the rotational axes of the tibiofemoral joint: a computational analysis.

An accurate method to locate of the flexion-extension (F-E) axis and longitudinal rotation (LR) axis of the tibiofemoral joint is required to accurately characterize tibiofemoral kinematics. A method was recently developed to locate these axes using an instrumented spatial linkage (ISL) (2012, "On the Estimate of the Two Dominant Axes of the Knee Using an Instrumented Spatial Linkage," J. Appl....

متن کامل

Effect of segmental joint stiffness on tunnel lining internal forces under static conditions

According to the wide application of segmental lining in mechanized tunneling, recognizing the behavior of segmental lining joints is important in tunnels designing. In the structural analysis of the tunnel segmental lining, segmental joints can be considered as elastic joints, and their stiffness characteristics are affected by the rotational, shear, and axial stiffness. The purpose of this wo...

متن کامل

On the estimate of the two dominant axes of the knee using an instrumented spatial linkage.

This article presents the validation of a technique to assess the appropriateness of a 2 degree-of-freedom model for the human knee, and, in which case, the dominant axes of flexion/extension and internal/external longitudinal rotation are estimated. The technique relies on the use of an instrumented spatial linkage for the accurate detection of passive knee kinematics, and it is based on the a...

متن کامل

Safety and Efficacy of Casting During COVID-19 Pandemic: A Comparison of the Mechanical Properties of Polymers Used for 3D Printing to Conventional Materials Used for the Generation of Orthopaedic Orthoses

To reduce the risk of spread of the novel coronavirus (COVID-19), the emerging protocols are advising for less physicianpatientcontact, shortening the contact time, and keeping a safe distance. It is recommended that unnecessary castingbe avoided in the events that alternative methods can be applied such as in stable ankle fractures, and hindfoot/midfoot/forefoot injuries. Fib...

متن کامل

Analysis of the Geometry of the Distal Femur and Proximal Tibia in the Osteoarthritic Knee: A 3D Reconstruction CT Scan Based Study of 449 Cases

    Background: The aim of this study is to evaluate the geometry of the distal femur and the proximal tibia in the osteoarthritic knee using 3D reconstructive CT scan imaging. Methods: 449 patients with knee osteoarthritis were treated surgically in our center with patient-specific technology total knee arthroplasty. Preoperatively, all the patients underwent a CT scan according to a standard ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of biomechanical engineering

دوره 136 1  شماره 

صفحات  -

تاریخ انتشار 2014