Patient and surgeon radiation exposure: comparison of standard and mini-C-arm fluoroscopy.
نویسندگان
چکیده
BACKGROUND Use of c-arm fluoroscopy is common in the operating room, outpatient clinic, and emergency department. Consequently, there is a concern regarding radiation exposure. Mini-c-arm fluoroscopes have gained popularity; however, few studies have quantified exposure during mini-c-arm imaging of a body part larger than a hand or wrist. The purpose of this study was to measure radiation exposure sustained by the patient and surgeon during the use of large and mini-c-arm fluoroscopy of an ankle specimen. METHODS Standard and mini-c-arm fluoroscopes were used to image a cadaver ankle specimen, which was suspended on an adjustable platform. Dosimeters were mounted at specific positions and angulations to detect direct and scatter radiation. Testing was conducted under various scenarios that altered the proximity of the specimen and the radiation source. We attempted to capture a range of exposure data under conditions ranging from a best to a worst-case scenario, as one may encounter in a procedural setting. RESULTS With all configurations tested, measurable exposure during use of the large-c-arm fluoroscope was considerably higher than that during use of the mini-c-arm fluoroscope. Patient and surgeon exposure was notably amplified when the specimen was positioned closer to the x-ray source. The exposure values that we measured during ankle fluoroscopy were consistently higher than the exposure values that have been recorded previously during hand or wrist imaging. CONCLUSIONS Exposure of the patient and surgeon to radiation depends on the tissue density and the shape of the imaged extremity. Elevated exposure levels can be expected when larger body parts are imaged or when the extremity is positioned closer to the x-ray source. When it is possible to satisfactorily image an extremity with use of the mini c-arm, it should be chosen over its larger counterpart.
منابع مشابه
Exposure to direct and scatter radiation with use of mini-C-arm fluoroscopy.
BACKGROUND Mini-c-arm fluoroscopy has become an important resource to the orthopaedic surgeon. Exposure of the orthopaedic surgical team to radiation during standard large-c-arm fluoroscopy has been well studied; however, little is known about the amount of exposure to which a surgical team is subjected with the use of mini-c-arm fluoroscopy. Moreover, there is controversy regarding the use of ...
متن کاملFluoroscopic Exposure With Use of Mini-C-Arm During Routine Hand Surgery: A Prospective Comparison of Hand Versus Eye Radiation Dosage.
Radiation exposure with use of intraoperative fluoroscopy is a potential orthopedic occupational risk factor. The purpose of this study was to perform a prospective comparison of hand versus eye radiation exposure associated with mini-C-arm utilization and to test the hypothesis that routine mini-C-arm does not yield hand or eye dosages exceeding current recommended levels. Over a 12-month peri...
متن کاملRadiation exposure during percutaneous nephrolithotomy; Is there a risk of the patient and the operating team?
Background: The aim of this study was to calculate the radiation amount exposed during percutaneous nephrolithototomy (PCNL) and to make the urologists and other staff sensitive about the radiation risk they were exposed to. Materials and Methods: We measured the radiation exposure during 114 cases of PCNL performed. Thermoluminescent dosimeters (TLD) were placed between the operation table and...
متن کاملHand and body radiation exposure with the use of mini C-arm fluoroscopy.
PURPOSE To determine whole body and hand radiation exposure to the hand surgeon wearing a lead apron during routine intraoperative use of the mini C-arm fluoroscope. METHODS Four surgeons (3 hand attending surgeons and 1 hand fellow) monitored their radiation exposure for a total of 200 consecutive cases (50 cases per surgeon) requiring mini C-arm fluoroscopy. Each surgeon measured radiation ...
متن کاملOccupational radiation exposure to the surgeon.
Increased use of intraoperative fluoroscopy exposes the surgeon to significant amounts of radiation. The average yearly exposure of the public to ionizing radiation is 360 millirems (mrem), of which 300 mrem is from background radiation and 60 mrem from diagnostic radiographs. A chest radiograph exposes the patient to approximately 25 mrem and a hip radiograph to 500 mrem. A regular C-arm expos...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Journal of bone and joint surgery. American volume
دوره 91 8 شماره
صفحات -
تاریخ انتشار 2009