Coronary excimer laser angioplasty: reduced complications and indium-111 platelet accumulation compared with thermal laser angioplasty.

نویسندگان

  • T A Sanborn
  • D Alexopoulos
  • J D Marmur
  • H Kahn
  • J J Badimon
  • L Badimon
  • V Fuster
چکیده

The relative safety and thrombogenicity of pulsed excimer and thermal laser angioplasty systems were compared in 20 normal coronary artery segments in a total of seven pigs. Using similar over the wire catheter systems and laser delivery periods of 3 to 5 s, thermal laser angioplasty was achieved with a 1.3 mm metal probe heated with 10 W of continuous argon laser energy and excimer laser angioplasty was performed with a 4.5F excimer laser catheter consisting of 13 concentrically arranged 200 microns fiber optics delivering 35 to 40 mJ/mm2 of xenon chloride (308 nm) excimer laser irradiation at a repetition rate of 25 to 30 Hz and a pulse duration of 120 ns. On angiography, the incidence of vessel perforation (1 in 10 versus 3 in 10) and abrupt vessel closure (0 in 10 versus 2 in 10) was less with excimer compared with thermal laser angioplasty. Macroscopically, there was a greater incidence of mural and occlusive thrombus formation after thermal laser than after pulsed excimer laser angioplasty. Histologic examination confirmed that this thrombogenicity was associated with greater charring and coagulation necrosis of the media. Quantitative indium-111-labeled platelet deposition was significantly increased after thermal laser angioplasty (median 87.2 x 10(6)/cm length) compared with excimer-treated (0.4 x 10(6)/cm length) or control (1.2 x 10(6)cm length) segments (p less than 0.001). Thus, excimer laser angioplasty was found to result in fewer complications and, as a consequence, less thrombosis and platelet accumulation than did thermal laser angioplasty.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Acute biological response to laser balloon angioplasty in the atherosclerotic rabbit.

Laser balloon angioplasty with Nd:YAG energy has been proposed as a method to seal intimal dissection and prevent elastic recoil after balloon angioplasty. To better define the vessel response to laser balloon angioplasty, its effects on luminal diameter, Indium-111 labelled platelet deposition, and histology were studied in 10 atherosclerotic rabbits. Balloon angioplasty was performed in both ...

متن کامل

Immediate and late outcome of excimer laser and balloon coronary angioplasty: a quantitative angiographic comparison based on matched lesions.

OBJECTIVES This study sought to compare acute lumen changes and late lumen narrowing during and after excimer laser-assisted balloon angioplasty, measured by quantitative coronary angiography, with the immediate and long-term outcome of balloon angioplasty alone. BACKGROUND Although excimer laser coronary angioplasty is used as an adjunct or alternative to balloon angioplasty, limited compara...

متن کامل

Randomised trial of excimer laser angioplasty versus balloon angioplasty for treatment of obstructive coronary artery disease.

BACKGROUND Excimer laser coronary angioplasty is reported to give excellent procedural results for treatment of complex coronary lesions, but this method has not been compared with balloon angioplasty in a randomised trial. METHODS Patients (n = 308) with stable angina and coronary lesions longer than 10 mm on visual assessment were included. 151 patients (158 lesions) were assigned randomly ...

متن کامل

Quantitative angiographic comparison of elastic recoil after coronary excimer laser-assisted balloon angioplasty and balloon angioplasty alone.

OBJECTIVES Coronary lumen changes during and after excimer laser-assisted balloon angioplasty were measured by quantitative coronary angiography, and the results were compared with the effects of balloon angioplasty alone. BACKGROUND Reduction of atherosclerotic tissue mass by laser ablation in the treatment of coronary artery disease may be more effective in enlarging the lumen than balloon ...

متن کامل

Morphological predictors of acute complications after percutaneous excimer laser coronary angioplasty. Results of a comprehensive angiographic analysis: importance of the eccentricity index.

BACKGROUND Percutaneous excimer laser coronary angioplasty (ELCA) is a new technique for recanalization of arteries obstructed by coronary atherosclerosis. This study was conducted to assess the complication rate and determine the influence of clinical and angiographic characteristics on complications after ELCA. METHODS AND RESULTS A detailed, quantitative, angiographic core laboratory analy...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of the American College of Cardiology

دوره 16 2  شماره 

صفحات  -

تاریخ انتشار 1990