Comparison of left ventricular ejection fraction and volumes in heart failure by echocardiography, radionuclide ventriculography and cardiovascular magnetic resonance; are they interchangeable?

نویسندگان

  • N G Bellenger
  • M I Burgess
  • S G Ray
  • A Lahiri
  • A J Coats
  • J G Cleland
  • D J Pennell
چکیده

AIMS To prospectively compare the agreement of left ventricular volumes and ejection fraction by M-mode echocardiography (echo), 2D echo, radionuclide ventriculography and cardiovascular magnetic resonance performed in patients with chronic stable heart failure. It is important to know whether the results of each technique are interchangable, and thereby how the results of large studies in heart failure utilizing one technique can be applied using another. Some studies have compared cardiovascular magnetic resonance with echo or radionuclude ventriculography but few contain patients with heart failure and none have compared these techniques with the current fast breath-hold acquisition cardiovascular magnetic resonance. METHODS AND RESULTS Fifty two patients with chronic stable heart failure taking part in the CHRISTMAS Study, underwent M-mode echo, 2D echo, radionuclude ventriculography and cardiovascular magnetic resonance within 4 weeks. The scans were analysed independently in blinded fashion by a single investigator at three core laboratories. Of the echocardiograms, 86% had sufficient image quality to obtain left ventricular ejection fraction by M-mode method, but only 69% by 2D Simpson's biplane analysis. All 52 patients tolerated the radionuclude ventriculography and cardiovascular magnetic resonance, and all these scans were analysable. The mean left ventricular ejection fraction by M-mode cube method was 39+/-16% and 29+/-15% by Teichholz M-mode method. The mean left ventricular ejection fraction by 2D echo Simpson's biplane was 31+/-10%, by radionuclude ventriculography was 24+/-9% and by cardiovascular magnetic resonance was 30+/-11. All the mean left ventricular ejection fractions by each technique were significantly different from all other techniques (P<0.001), except for cardiovascular magnetic resonance ejection fraction and 2D echo ejection fraction by Simpson's rule (P=0.23). The Bland-Altman limits of agreement encompassing four standard deviations was widest for both cardiovascular magnetic resonance vs cube M-mode echo and cardiovascular magnetic resonance vs Teichholz M-mode echo at 66% each, and was 58% for radionuclude ventriculography vs cube M-mode echo, 44% for cardiovascular magnetic resonance vs Simpson's 2D echo, 39% for radionuclide ventriculography vs Simpson's 2D echo, and smallest at 31% for cardiovascular magnetic resonance-radionuclide ventriculography. Similarly, the end-diastolic volume and end-systolic volume by 2D echo and cardiovascular magnetic resonance revealed wide limits of agreement (52 ml to 216 ml and 11 ml to 188 ml, respectively). CONCLUSION These results suggest that ejection fraction measurements by various techniques are not interchangeable. The conclusions and recommendations of research studies in heart failure should therefore be interpreted in the context of locally available techniques. In addition, there are very wide variances in volumes and ejection fraction between techniques, which are most marked in comparisons using echocardiography. This suggests that cardiovascular magnetic resonance is the preferred technique for volume and ejection fraction estimation in heart failure patients, because of its 3D approach for non-symmetric ventricles and superior image quality.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Comparison of techniques for the measurement of left ventricular function following cardiac transplantation.

BACKGROUND Assessment of graft function after cardiac transplantation is essential for patient management and clinical research. Previous studies have found that the left ventricular (LV) ejection fraction (EF) by echocardiography (echo), radionuclide ventriculography (RNV), and cardiovascular magnetic resonance (CMR) is discrepant in patients with heart failure. METHOD Twelve patients underw...

متن کامل

The Correlation between Left and Right Ventricular Ejection Fractions in Patients with Ischemic Heart Disease, Documented by Cardiac Magnetic Resonance Imaging

Introduction: The correlation between right and left ventricular ejection fractions (RVEF and LVEF, respectively) has been studied in only a small number of patients with a marked decrease in RVEF and LVEF. The aim of the present study was to compare LVEF and RVEF in patients with ischemic heart disease. RVEF and LVEF were measured by Cardiovascular Magnetic Resonance (CMR) imaging. Materials a...

متن کامل

Unreliability of cardiothoracic ratio as a marker of left ventricular impairment: comparison with radionuclide ventriculography and echocardiography.

BACKGROUND The size of the heart assessed by cardiothoracic ratio on chest radiography is often used as a screening test for the presence of heart failure and for assessing its severity. METHODS We compared cardiothoracic ratio (CTR), left ventricular ejection fraction (LVEF) from radionuclide ventriculography, and left ventricular dimensions from echocardiography in a population of 91 patien...

متن کامل

Left ventricular ejection fraction calculation from automatically selected and processed diastolic and systolic frames in short-axis cine-MRI.

The calculation of the left ventricular ejection fraction (LVEF) is dependent upon the accurate measurement of diastolic and systolic left ventricular volumes. Although breath-hold cine magnetic resonance imaging (MRI) allows coverage of the whole cardiac cycle with an excellent time resolution, many authors rely on the visual selection of diastolic and the systolic short-axis slices in order t...

متن کامل

The effect of ventricular pacing on measurements of left ventricular function: a comparison between echocardiographic methods and with radionuclide ventriculography.

AIMS Different methods exist for measuring left ventricular function echocardiographically; each may be error prone due to the abnormal pattern of ventricular activation during pacing. METHODS AND RESULTS Echocardiography was undertaken on 307 patients with permanent pacemakers; a subset of 57 underwent radionuclide ventriculography. Intrinsic and paced beats were analysed for left ventricula...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • European heart journal

دوره 21 16  شماره 

صفحات  -

تاریخ انتشار 2000