Advances in tribological testing of artificial joint biomaterials using multidirectional pin-on-disk testers.
نویسندگان
چکیده
The introduction of numerous formulations of Ultra-high molecular weight polyethylene (UHMWPE), which is widely used as a bearing material in orthopedic implants, necessitated screening of bearing couples to identify promising iterations for expensive joint simulations. Pin-on-disk (POD) testers capable of multidirectional sliding can correctly rank formulations of UHMWPE with respect to their predictive in vivo wear behavior. However, there are still uncertainties regarding POD test parameters for facilitating clinically relevant wear mechanisms of UHMWPE. Studies on the development of POD testing were briefly summarized. We systematically reviewed wear rate data of UHMWPE generated by POD testers. To determine if POD testing was capable of correctly ranking bearings and if test parameters outlined in ASTM F732 enabled differentiation between wear behavior of various formulations, mean wear rates of non-irradiated, conventional (25-50kGy) and highly crosslinked (≥90kGy) UHMWPE were grouped and compared. The mean wear rates of non-irradiated, conventional and highly crosslinked UHMWPEs were 7.03, 5.39 and 0.67mm(3)/MC. Based on studies that complied with the guidelines of ASTM F732, the mean wear rates of non-irradiated, conventional and highly crosslinked UHMWPEs were 0.32, 0.21 and 0.04mm(3)/km, respectively. In both sets of results, the mean wear rate of highly crosslinked UHMPWE was smaller than both conventional and non-irradiated UHMWPEs (p<0.05). Thus, POD testers can compare highly crosslinked and conventional UHMWPEs despite different test parameters. Narrowing the allowable range for standardized test parameters could improve sensitivity of multi-axial testers in correctly ranking materials.
منابع مشابه
Storing Period Effects on Wear Properties in Uhmwpe Inserts for Artificial Joints
INTRODUCTION Artificial joints have been increasingly used in treatment of physically disabled people who have severely damaged joints. In artificial joint, ultra high molecular weight polyethylene (UHMWPE) has been usually employed as a buffer because of its very low friction coefficient, high wear resistance, high impact strength, and biocompatibility. [1]. When artificial joints were replace...
متن کاملCan Pin-on-Disk Testing Be Used to Assess the Wear Performance of Retrieved UHMWPE Components for Total Joint Arthroplasty?
The objective of this study was to assess the suitability of using multidirectional pin-on-disk (POD) testing to characterize wear behavior of retrieved ultrahigh molecular weight polyethylene (UHMWPE). The POD wear behavior of 25 UHMWPE components, retrieved after 10 years in vivo, was compared with 25 that were shelf aged for 10-15 years in their original packaging. Components were gamma ster...
متن کاملEffects of Select Fluids on the Friction of Metal-on-Polyethylene
Lubricants are important factors in the tribology of total joint arthroplasty (TJA) surfaces, which are primarily comprised of a polished metallic or ceramic component articulating on an Ultra-High Molecular Weight Polyethylene (PE) surface. Wear particles released from the PE surface are the primary cause of TJA failure. The human body responds to the foreign, microscale particles by activatin...
متن کاملEffect of Deposition Regime on Tribological Behavior of VK8Coating on 1.2344 Tool Steel Obtained by Electro Spark Method
In this study, the effect of deposition regime on the wear behavior of VK8 coating applied to 1.2344 tool steel by electro spark method (ESA) was investigated. After sample preparation, the coating with high and low regime was applied. To assess the wear behavior, the pin on disk test method according to ASTM G99-04, was used. For surveying surface morphology and cross section of the coating as...
متن کاملTribological performance of the biological components of synovial fluid in artificial joint implants
The concentration of biological components of synovial fluid (such as albumin, globulin, hyaluronic acid, and lubricin) varies between healthy persons and osteoarthritis (OA) patients. The aim of the present study is to compare the effects of such variation on tribological performance in a simulated hip joint model. The study was carried out experimentally by utilizing a pin-on-disk simulator o...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of the mechanical behavior of biomedical materials
دوره 31 شماره
صفحات -
تاریخ انتشار 2014