Polyethylene wear in cemented metal-backed acetabular cups.

نویسندگان

  • H E Cates
  • P M Faris
  • E M Keating
  • M A Ritter
چکیده

We examined radiographic polyethylene wear in 233 cemented total hip arthroplasties (201 patients) with either a metal-backed or a non-metal-backed acetabular cup. All patients had identical cemented one-piece titanium femoral stems with a femoral head diameter of 28 mm. The mean linear wear rate was 0.11 mm/yr in metal-backed sockets and 0.08 mm/yr in non-metal-backed sockets (p = 0.0002). The mean volumetric wear rate was 66.2 mm3/yr in the metal-backed sockets and 48.2 mm3/yr in the polyethylene sockets (p = 0.0002). The addition of metal backing to a cemented acetabular cup therefore resulted in a 37% increase in mean polyethylene wear rates which may partially explain the higher failure rate of cemented metal-backed cups. Linear regression analysis also implicated increased follow-up time (log), gross acetabular migration, metal backing and male gender in increasing polyethylene wear. We advocate the use of an all-polyethylene cup in cemented total hip arthroplasty. The increased polyethylene wear must also cause concern about the wear rate of uncemented metal-backed acetabular sockets.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Polyethylene wear in total hip arthroplasty.

Polyethylene wear increasingly has been blamed for osteolysis and granuloma formation after total joint arthroplasty. The authors evaluated the wear difference between bearing surfaces of stainless steel, cobalt chrome, and titanium alloy. They also compared cemented all-polyethylene molded cups with cemented metal-backed molded acetabular cups. These were compared with molded and machined poly...

متن کامل

Polyethylene wear in uncemented acetabular components.

We measured polyethylene wear in 231 porous-coated uncemented acetabular cups. We divided the hips into two groups according to the fixation of the femoral component, by cementing (n = 97) or press-fit (n = 134). Follow-up was from three to five years. The patients in two sub-groups were matched for weight, diagnosis, sex, age and length of follow-up. The linear wear rate of cups articulated wi...

متن کامل

Finite Element Analysis Stress Distribution Acetabular Cup Biomaterial FINITE ELEMENT ANALYSIS OF PERI-ACETABULAR STRESS OF CEMENTED, METAL-BACKED, AND POROUS TANTALUM BACKED ACETABULAR COMPONENTS

Introduction: In contemporary THA there are two broad approaches to acetabular cup design and materials. One is an all-polyethylene design with polymethylmethacrylate (PMMA) cement. PMMA possesses a modulus of elasticity of about 2 GPa, which is between that of trabecular and subchondral bone. The second more popular approach, especially for active, younger patients, is the porous metal-backed ...

متن کامل

Total hip arthroplasty in patients less than 40 years old with avascular necrosis of the femoral head. A 5 to 19-year follow-up study.

We evaluated 34 total hip arthroplasties (THA) for avascular necrosis of the femoral head in 26 patients who were younger than 40 years at the time of the index operation. The average age at the time of THA was 283 (17-38) years. The mean follow-up was 10.9 (5-19) years. We used 6 cemented and 28 cementless acetabular components and a cemented polished tapered stem in all hips. The mean Charnle...

متن کامل

Fixation of ultrahigh-molecular-weight polyethylene liners to metal-backed acetabular cups.

Locking mechanisms and metal-liner interface surfaces of six modular acetabular systems were evaluated to determine their effect on micromotion and backside wear of the polyethylene liner. Rotational and axial motion between the metal shell and polyethylene liner was measured in the Duraloc (DePuy, Warsaw, IN), Harris-Galante (Zimmer, Warsaw, IN), Impact (Biomet, Warsaw, IN), Lip Loc (Biomet), ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Journal of bone and joint surgery. British volume

دوره 75 2  شماره 

صفحات  -

تاریخ انتشار 1993