A multicentre evaluation of the Boehringer Mannheim/Hitachi 704 analysis system.

نویسندگان

  • P M Bayer
  • M Knedel
  • N Montalbetti
  • S Brenna
  • L Prencipe
  • A Vassault
  • M Bailly
  • H T Phung
  • W Bablok
  • W Poppe
چکیده

The selective multitest Boehringer Mannheim/Hitachi 704 analysis system was examined according to the ECCLS guidelines in a multicentre evaluation involving four laboratories. Ten routine parameters, covering most of the application settings of the instrument, were measured in the respective laboratory at temperatures 25, 30 or 37 degrees C. The trial lasted four months and gave more than 40,000 data. It yielded the following results: 1. Within the four laboratories the mean coefficients of variation for three control sera at different concentrations were found to be equal to or better than 1.6% for the within-run imprecision and 2.8% or better for the between-day imprecision. 2. No drift was observed during eight hours. 3. Because of the high linear measuring range a re-run analysis was seldom necessary. 4. Sample-related carry-over was not seen. Reagent-dependent carry-over was measured from cholesterol to uric acid and from triacylglycerols to lipase. Through modification of the cholesterol and triacylglycerol reagents, the carry-over effect was practically eliminated. 5. The recovery of the assigned values of control sera showed average values between 99 and 104%. For bilirubin, creatinine, creatine kinase and alanine aminotransferase some control sera showed deviations greater than 10%. 6. In all cases, regression analysis of the results obtained in comparisons of the present instrument with the Hitachi 705 or 737 yielded slopes close to unity with extreme values of 0.95 and 1.06. 7. During the entire evaluation period there was no malfunction or breakdown of the instruments. The evaluators came to the conclusion that the analytical performance as well as the reliability and practicability of the Hitachi 704 can be rated as excellent.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Multicentre evaluation of the Boehringer Mannheim/Hitachi 911 Analysis System

The analytical performance and practicability of the Boehringer Mannheim (BM)/Hitachi 911 analysis system have been assessed in a multicentre evaluation, which involved six laboratories from European countries. Analytes commonly used in classical clinical chemistry were tested in a core programme, which mainly followed the ECCLS guidelines. In addition, a satellite programme covered other analy...

متن کامل

Multicentre evaluation of the Boehringer Mannheim/Hitachi 917 analysis system

The new selective access analysis system BM/Hitachi 917 was evaluated in an international multicentre study, mainly according to the ECCLS protocol for the evaluation of analysers in clinical chemistry. Forty-three different analytes, covering 56 different methods--enzymes, substrates, electrolytes, specific proteins, drugs and urine applications--were tested in seven European clinical chemistr...

متن کامل

Standardization of a commercial (Boehringer Mannheim diagnostics) enzymic method for cholesterol.

The Laboratory Standardization Panel of the National Cholesterol Education Program recommends that cholesterol method accuracy ideally be within 3% of the true value determined by the Abell-Kendall Reference Method, a component of the National Reference System for Cholesterol. As one of the Abell-Kendall network laboratories established to facilitate cholesterol standardization, the approach we...

متن کامل

Evaluation of the Hitachi 717 analyser

The selective multitest Boehringer Mannheim Hitachi 717 analyser was evaluated according to the guidelines of the Comisión de Instrumentación de la Sociedad Española de Química Clinica and the European Committee for Clinical Laboratory Standards. The evaluation was performed in two steps: examination of the analytical units and evaluation in routine operation.THE EVALUATION OF THE ANALYTICAL UN...

متن کامل

Comparison of a commercial method for total protein with a candidate reference method.

The biuret method for total protein has been compared on the Boehringer Mannheim Diagnostics (BMD) Hitachi 717 Analyzer with a candidate reference method in efforts to standardize the BMD method. The methods were compared using 115 paired serum specimens collected during morning rounds at the Roger Williams Hospital. Results from the test method were compared to the reference method using a sta...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of clinical chemistry and clinical biochemistry. Zeitschrift fur klinische Chemie und klinische Biochemie

دوره 25 12  شماره 

صفحات  -

تاریخ انتشار 1987