[Evaluation of combination of BACTEC mycobacteria growth indicator tube 960 system and Ogawa media for mycobacterial culture].
نویسندگان
چکیده
BACKGROUND The combined use of liquid media and solid media is recommended for mycobacterial culture. We evaluated diagnostic performance of combination of BACTEC Mycobacteria Growth Indicator Tube (MGIT; Becton Dickinson, USA) and 2% Ogawa media (Korean Institute of Tuberculosis, Korea) for recovery of mycobacteria. METHODS In September 2007, 1,764 specimens from 1,059 patients were cultured with MGIT and Ogawa. Acid fast bacilli (AFB) smear was fluorochrome-stained. The isolates were identified into Mycobacterium tuberculosis (MTB) and nontuberculous mycobacteria (NTM) with PCR using Seeplex TB Detection Kit (Seegene, Korea). Recovery rate, time to detection (TTD), contamination rate, mixed growth rate and species distribution were analyzed. RESULTS Two hundred thirty-five specimens (13.3%) from 165 patients (15.6%) were positive for mycobacterial culture. Recovery rates of mycobacteria from the group using both media, MGIT only, and Ogawa only were 13.3%, 12.1%, and 7.8%, respectively. While MGIT recovered 98.9% of MTB and 79.7% of NTM, Ogawa recovered 65.9% of MTB and 54.1% of NTM. TTDs of total mycobacteria/MTB/NTM in MGIT and Ogawa were 10.6/11.4/9.7 days and 31/29/33 days, respectively. MGIT TTDs of total mycobacteria/MTB/NTM from AFB-positive specimens were significantly shorter than those of AFB-negative specimens; 8.2/9.5/4.4 days vs 11.6/12.7/10.7 days. Contamination and mixed growth rate of MGIT were 9.6% and 3.7%. Primary culture of Ogawa recovered 1 MTB and 1 NTM among the 170 MGIT-contaminated specimens and 38 mycobacteria among 66 specimens that showed mixed cultures of MGIT. CONCLUSIONS MGIT warrants sensitive and rapid isolation of mycobacteria. However, the combination of MGIT and Ogawa is more desirable to recover mycobacteria in the case of contaminations or mixed cultures.
منابع مشابه
Comparison of Ogawa Media, BACTEC MGIT 960 System and TB/NTM Real-Time PCR for Detecting Mycobacterium Species
Background: Mycobacterial infection is a problem throughout the world along with the increase of immunocompromised patients. For this reason, there have been many methods for faster and more accurate diagnosis. In this study, we evaluated several laboratory methods for mycobacterial infection. Methods: From January to December 2009, 635 specimens were cultured with mycobacteria growth indicator...
متن کاملUse of the BACTEC Mycobacteria Growth Indicator Tube 960 automated system for recovery of Mycobacteria from 9,558 extrapulmonary specimens, including urine samples.
The BACTEC Mycobacteria Growth Indicator Tube 960 (MGIT 960) system was applied for recovery of mycobacteria from extrapulmonary specimens and compared with solid media (Löwenstein-Jensen and Stonebrink). A total of 9,558 specimens were investigated, comprising 3,074 body fluids, 1,878 tissues, and 2,069 urine samples, from which the recovery of mycobacteria was not yet established for MGIT 960...
متن کاملComparison of the BACTEC MGIT 960 and BACTEC 460TB systems for detection of mycobacteria in clinical specimens.
The reliability of the Mycobacteria Growth Indicator Tube (MGIT) 960 system for rapid detection of mycobacteria in clinical specimens was evaluated and compared to the radiometric method (BACTEC 460TB) and to mycobacterial culture on Lowenstein-Jensen (LJ) medium. Clinical specimens (n = 590) were tested without selection. A total of 121 (20.5%) isolates of mycobacteria were recovered; 98 (81.0...
متن کاملGrowth detection failures by the nonradiometric Bactec MGIT 960 mycobacterial culture system.
Mycobacterial growth in liquid culture can go undetected by automated, nonradiometric growth detection systems. In our laboratory, instrument-negative tubes from the Bactec MGIT 960 system are inspected visually for clumps suggestive of mycobacterial growth, which (if present) are examined by acid-fast smear analysis. A 3-year review demonstrated that ∼1% of instrument-negative MGIT cultures co...
متن کاملImpact of Implementation of an Automated Liquid Culture System on Diagnosis of Tuberculous Pleurisy
This study was conducted to evaluate the impact of implementation of an automated liquid culture system on the diagnosis of tuberculous pleurisy in an HIV-uninfected patient population. We retrospectively compared the culture yield, time to positivity, and contamination rate of pleural effusion samples in the BACTEC Mycobacteria Growth Indicator Tube 960 (MGIT) and Ogawa media among patients wi...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- The Korean journal of laboratory medicine
دوره 28 4 شماره
صفحات -
تاریخ انتشار 2008