Biomechanical In Vitro - Stability Testing on Human Specimens of a Locking Plate System Against Conventional Screw Fixation of a Proximal First Metatarsal Lateral Displacement Osteotomy
نویسندگان
چکیده
INTRODUCTION The aim of this study was to examine resistance to angulation and displacement of the internal fixation of a proximal first metatarsal lateral displacement osteotomy, using a locking plate system compared with a conventional crossed screw fixation. MATERIALS AND METHODOLOGY Seven anatomical human specimens were tested. Each specimen was tested with a locking screw plate as well as a crossed cancellous srew fixation. The statistical analysis was performed by the Friedman test. The level of significance was p = 0.05. RESULTS We found larger stability about all three axes of movement analyzed for the PLATE than the crossed screws osteosynthesis (CSO). The Friedman test showed statistical significance at a level of p = 0.05 for all groups and both translational and rotational movements. CONCLUSION The results of our study confirm that the fixation of the lateral proximal first metatarsal displacement osteotomy with a locking plate fixation is a technically simple procedure of superior stability.
منابع مشابه
Biomechanical Comparison Study of Three Fixation Methods for Proximal Chevron Osteotomy of the First Metatarsal in Hallux Valgus
Background Fixation of proximal chevron metatarsal osteotomy has been accomplished using K-wires traditionally and with a locking plate recently. However, both methods have many disadvantages. Hence, we developed an intramedullary fixation technique using headless cannulated screws and conducted a biomechanical study to evaluate the superiority of the technique to K-wire and locking plate fixat...
متن کاملBiomechanical Comparison of Two Ludloff Metatarsal Osteotomy Fixations: Compression Screws vs. Locking Plate
BACKGROUND: Ludloff osteotomy fixation with compression screws has been widely used in the treatment of Hallux Valgus deformities [1, 2]. Generally six weeks of non-weight-bearing is required postoperatively to allow bone union. Besides the significant inconvenience due to non-weight-bearing, noncompliance may induce malunion, nonunion and loss of correction, indicating screw fixation is not st...
متن کاملUnstable proximal extraarticular tibia fractures: a biomechanical evaluation of four methods of fixation.
OBJECTIVE To compare the biomechanical stability of extraarticular proximal tibia fractures reconstructed using a double-plate construct, locking plate system, hybrid external fixator, and single lateral periarticular plate, all from the same manufacturer. DESIGN Standardized proximal tibial fractures (AO classification 41-A3.2 and A3.3) in synthetic tibiae were stabilized using one of the fo...
متن کاملComparison of lateral locking plate and antiglide plate for fixation of distal fibular fractures in osteoporotic bone: a biomechanical study.
OBJECTIVES The purpose of this study was to compare the biomechanical properties of posterolateral antiglide plating and lateral locked plating for fixation of displaced short oblique fractures of the fibula in osteoporotic bone. METHODS Short oblique fractures of the distal fibula at the level of the syndesmosis were simulated with a fibular osteotomy and ligamentous sectioning in 18 paired ...
متن کاملBiomechanical Properties of a New Anatomical Locking Metal Block Plate for Opening Wedge High Tibial Osteotomy: Uniplane Osteotomy
PURPOSE The purpose of this study was to evaluate the biomechanical properties of a new anatomical locking metal block plate by comparing the initial biomechanical stability of three different fixation constructs for open wedge high tibial osteotomy (HTO). MATERIALS AND METHODS Sawbones composite tibiae were used to make a 10-mm opening osteotomy model with uniplane technique. The osteotomy w...