Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification

نویسندگان

  • Valeria Casanova Borca
  • Massimo Pasquino
  • Giuliana Russo
  • Pierangelo Grosso
  • Domenico Cante
  • Piera Sciacero
  • Giuseppe Girelli
  • Maria Rosa La Porta
  • Santi Tofani
چکیده

Radiochromic film has become an important tool to verify dose distributions in highly conformal radiation therapy such as IMRT. Recently, a new generation of these films, EBT3, has become available. EBT3 has the same composition and thickness of the sensitive layer of the previous EBT2 films, but its symmetric layer configuration allows the user to eliminate side orientation dependence, which is reported for EBT2 films. The most important EBT3 characteristics have been investigated, such as response at high-dose levels, sensitivity to scanner orientation and postirradiation coloration, energy and dose rate dependence, and orientation dependence with respect to film side. Additionally, different IMRT fields were measured with both EBT3 and EBT2 films and evaluated using gamma index analysis. The results obtained show that most of the characteristics of EBT3 film are similar to the EBT2 film, but the orientation dependence with respect to film side is completely eliminated in EBT3 films. The study confirms that EBT3 film can be used for clinical practice in the same way as the previous EBT2 film.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

A Feasibility Study of IMRT of Lung Cancer Using Gafchromic EBT3 Film

Background: Intensity modulated radiation therapy (IMRT) is an advanced method for delivery of three dimensional therapies, which provides optimal dose distribution with giving multiple nonuniform fluency to the patient. The complex dose distribution of IMRT should be checked to ensure that the accurate dose is delivered. Today, film dosimetry is a powerful tool for radiotherapy treatment Quali...

متن کامل

The use of EBT3 film and Delta4 for the dosimetric verification of EclipseTM treatment planning system in a heterogeneous chest phantom: an IMRT technique

Background: This study aimed to evaluate the dose calculation accuracy of EclipseTM treatment planning system (TPS) in a heterogeneous chest phantom with the intensity modulated radiotherapy (IMRT) technique using EBT3 film and Delta4. Materials and Methods: Two IMRT plans (A and B) were prepared for radiotherapy of the heterogeneous chest phantom. Plan A was between the left lung and the surro...

متن کامل

Sum signal dosimetry: A new approach for high dose quality assurance with Gafchromic EBT3

Gafchromic EBT3 film dosimetry in radiosurgery (RS) and hypofractionated radiotherapy (HRT) is complicated by the limited film accuracy at high fractional doses. The aim of this study is to develop and evaluate sum signal (SS) film dosimetry to increase dose resolution at high fractional doses, thus allowing for use of EBT3 for dose distribution verification of RS/HRT treatments. To characteriz...

متن کامل

The combination of the error correction methods of GAFCHROMIC EBT3 film

PURPOSE The aim of this study was to combine a set of methods for use of radiochromic film dosimetry, including calibration, correction for lateral effects and a proposed triple-channel analysis. These methods can be applied to GAFCHROMIC EBT3 film dosimetry for radiation field analysis and verification of IMRT plans. METHODS A single-film exposure was used to achieve dose calibration, and th...

متن کامل

SU-E-T-121: Dosimetric Characterization of Gafchromic Film EBT3 Using Vidar DosimetryPro Advantage RED and EPSON Expression 10000XL Scanners.

PURPOSE The purpose of this paper is to characterize EBT3 using two types of scanner, analyzing the factors of influence of each dosimetry system. METHODS The film used in this study was GAFCHROMIC EBT3, the films were exposed at a dose range between 0Gy a 9Gy in a solid water phantom, SSD=100cm, 5cm depth and perpendicularly to the 6MV photon beam generated by a Novalis TX linear accelerator...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 14  شماره 

صفحات  -

تاریخ انتشار 2013