A comparison between two different automated total 25-hydroxyvitamin D immunoassay methods using liquid chromatography-tandem mass spectrometry
نویسندگان
چکیده
INTRODUCTION Total 25-hydroxyvitamin D [25(OH)D] is the most reliable indicator of vitamin D status. In this study, we compared two automated immunoassay methods, the Abbott Architect 25-OH Vitamin D assay and the Roche Cobas Vitamin D total assay, with the liquid chromatography-tandem mass spectrometry (LC-MS/MS). MATERIALS AND METHODS One hundred venous blood samples were randomly selected from routine vitamin D tests. Two of the serum aliquots were analyzed at the Abbott Architect i2000 and the Roche Cobas 6000's module e601 in our laboratory within the same day. The other serum aliquots were analyzed at the LC-MS/MS in different laboratory. Passing-Bablok regression analysis and Bland-Altman plot were used to compare methods. Inter-rater agreement was analyzed using kappa (κ) analysis. RESULTS The Roche assay showed acceptable agreement with the LC-MS/MS based on Passing-Bablok analysis (intercept: -5.23 nmol/L, 95% CI: -8.73 to 0.19; slope: 0.97, 95% CI: 0.77 to 1.15). The Abbott assay showed proportional (slope: 0.77, 95% CI: 0.67 to 0.85) and constant differences (intercept: 17.08 nmol/L; 95% CI: 12.98 to 21.39). A mean bias of 15.1% was observed for the Abbott and a mean bias of -14.1% was observed for the Roche based on the Bland-Altman plots. We found strong to nearly perfect agreement in vitamin D status between the immunoassays and LC-MS/MS. (κ: 0.83 for Abbott, κ: 0.93 for Roche) using kappa analysis. CONCLUSION Both immunoassays demonstrated acceptable performance, but the Roche Cobas assay demonstrated better performance than the Abbott Architect in the studied samples.
منابع مشابه
Efficacy of the Measurement of 25-Hydroxyvitamin D2 and D3 Levels by Using PerkinElmer Liquid Chromatography-Tandem Mass Spectrometry Vitamin D Kit Compared With DiaSorin Radioimmunoassay Kit and Elecsys Vitamin D Total Assay
Vitamin D2 (ergocalciferol) and D3 (cholecalciferol) can be procured from exogenous sources. These are then metabolized to 25-hydroxyvitamin D (25OHD2 and 25OHD3) in the liver. Measuring the levels of both 25OHD2 and 25OHD3 is imperative in assessing clinical nutritional status [1]. Vitamin D2 or D3 is provided as a vitamin D supplement in many countries. Serum 25OHD levels can be measured by c...
متن کامل[Vitamin D measurement: pre-analytical and analytical considerations].
The constantly increasing requests for the measurement of serum 25-hydroxyvitamin D over the last years has led reagent manufacturers to market different automated and semi-automated methods, that being unfortunately not fully harmonized, yield different results. Liquid chromatography coupled to tandem mass spectrometry has more recently been introduced. This approach allows the distinction bet...
متن کاملMeasurement of circulating 25-hydroxyvitamin D: A historical review
The constantly increasing requests for the measurement of serum 25-hydroxyvitamin D over the last years has led reagent manufacturers to market different automated and semi-automated methods, that being unfortunately not fully harmonized, yield different results. Liquid chromatography coupled to tandem mass spectrometry (LC/MS2) has more recently been introduced. This approach allows the distin...
متن کاملPerformance evaluation of two immunoassays for 25-hydroxyvitamin D
Although immunoassays in measuring 25-hydroxyvitamin D [25(OH)D] have been improved recently, relatively large differences are still seen between results of 25(OH)D measured by immunoassays and by liquid chromatography-tandem mass spectrometry (LC-MS/MS). In the present studies, we compared two immunoassays with LC-MS/MS for measuring 25(OH)D concentrations. Concentrations of 25-hydroxyvitamin ...
متن کاملFour automated 25-OH total vitamin D immunoassays and commercial liquid chromatography tandem-mass spectrometry in Finnish population.
BACKGROUND Comparing four fully automated 25-OH-D immunoassays to a commercially available liquid chromatography-tandem mass spectrometry (LC-MS/MS) method with human serum samples from Finnish population. METHODS 400 samples were analyzed with the Liaison Total Vitamin D, the IDS-iSYS 25-Hydroxy Vitamin D, the ARCHITECT 25-OH Vitamin D, the ADVIA Centaur Vitamin D Total, and a commercially a...
متن کامل