International comparison of the illuminance responsivity scales and units of luminous flux maintained at the HUT (Finland) and the NIST (USA)

نویسنده

  • Y. Ohno
چکیده

An international comparison has been conducted to compare the illuminance responsivity scales (A/lx) and the units of luminous  ux (lm) maintained at the National Institute of Standards and Technology (NIST, USA) and the Helsinki University of Technology (HUT, Finland). Both laboratories realize the illuminance unit by absolutely calibrated photometers and the luminous  ux unit by the absolute integrating-sphere method. Standard photometers were used as transfer standards for the illuminance responsivity comparison, and standard lamps in the luminous  ux comparison. The ratio of the measured illuminance responsivity values (HUT/NIST) was 0.9992 with an expanded uncertainty of 0.0013, and the ratio of the measured luminous  ux values was 1.0006 with an expanded uncertainty of 0.0018. The relative expanded uncertainties of the agreement of the units, including the uncertainties of the realizations of the units as well as the uncertainty of the comparison, were 0.0047 and 0.0101, respectively.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Trilateral Intercomparison of Photometric Units Maintained at NIST (USA), NPL (UK), and PTB (Germany)

Physikalisch-Technische Bundesanstalt, Abteilung Optik, Photometrie, Postfach 33 45, 38023 Braunschweig, Germany A trilateral intercomparison of photometric units between NIST (USA), NPL (UK), and PTB (Germany) has been conducted to update the knowledge of the relationship between the photometric units disseminated in the three countries. The luminous intensity unit (cd), the luminous responsiv...

متن کامل

1993 Intercomparison of Photometric Units Maintained at NIST (USA) and PTB (Germany)

A bilateral intercomparison of photometric units between NIST, USA and PTB, Germany has been conducted to update the knowledge of the relationship between the photometric units disseminated in each country. The luminous intensity unit (cd) and the luminous flux unit (lm) maintained at both laboratories are compared by circulating transfer standard lamps. Also, the photometric responsivity sv is...

متن کامل

Realization of NIST 1995 Luminous Flux Scale using Integrating Sphere Method

A new NIST luminous flux scale has been realized using the integrating sphere method, on which theoretical and experimental studies were previously conducted and reported. The NIST 2 m integrating sphere was modified to add an opening and an additional baffle to apply this method. Flux from an external source (1000 W quartz halogen lamp) is introduced into the sphere through a limiting aperture...

متن کامل

Improved Photometric Standards and Calibration Procedures at NIST

NIST has recently established a detector-based luminous intensity unit (candela, cd), which is derived from the NIST absolute cryogenic radiometer. Subsequently, the luminous flux unit (lumen, lm) and the luminance unit (cd/m2) have been established based on the detector-based candela, and now all the NIST photometric units are tied to the cryogenic radiometer. The illuminance unit is realized ...

متن کامل

Realization of the unit of luminous flux at the HUT using the absolute integrating-sphere method

A description of a detector-based realization of the unit of luminous flux (lumen) at the Helsinki University of Technology (HUT) is presented. The realization is based on the absolute integrating-sphere method developed at the National Institute of Standards and Technology (NIST), with some modifications. The measurement set-up consists of a 1.65 m integrating sphere, two photometers, a precis...

متن کامل