Normative findings of electrically evoked compound action potential measurements using the neural response telemetry of the Nucleus CI24M cochlear implant system.

نویسندگان

  • D Cafarelli Dees
  • N Dillier
  • W K Lai
  • E von Wallenberg
  • B van Dijk
  • F Akdas
  • M Aksit
  • C Batman
  • A Beynon
  • S Burdo
  • J-M Chanal
  • L Collet
  • M Conway
  • C Coudert
  • L Craddock
  • H Cullington
  • N Deggouj
  • B Fraysse
  • S Grabel
  • J Kiefer
  • J G Kiss
  • T Lenarz
  • A Mair
  • S Maune
  • J Müller-Deile
  • J-P Piron
  • S Razza
  • C Tasche
  • H Thai-Van
  • F Toth
  • E Truy
  • A Uziel
  • G F Smoorenburg
چکیده

One hundred and forty-seven adult recipients of the Nucleus 24 cochlear implant system, from 13 different European countries, were tested using neural response telemetry to measure the electrically evoked compound action potential (ECAP), according to a standardised postoperative measurement procedure. Recordings were obtained in 96% of these subjects with this standardised procedure. The group results are presented in terms of peak amplitude and latency, slope of the amplitude growth function and ECAP threshold. The effects of aetiological factors and the duration of deafness on the ECAP were also studied. While large intersubject variability and intrasubject variability (across electrodes) were found, results fell within a consistent pattern and a normative range of peak amplitudes and latencies was established. The aetiological factors had little effect on the ECAP characteristics. However, age affected ECAP amplitude and slope of the amplitude growth function significantly; i.e., the amplitude is higher in the lowest age category (15-30 years). Principal component analysis of the ECAP thresholds shows that the thresholds across 5 electrodes can be described by two factors accounting for 92% of the total variance. The two factors represent the overall level of the threshold profiles ('shift') and their slopes across the electrode array ('tilt'). Correlation between these two factors and the same factors describing the T- and C-levels appeared to be moderate, in the range of 0.5-0.6.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Application of Auditory Evoked Potentials in Special Cases

Smoorenburg GF: Normative findings of electrically evoked compound action potential measurement using the neural response telemetry of the Nucleus CI24M cochlear implant system. of autonomic and sensory neuropathy and the impairment of visual and auditory evoked potentials in type-1 diabetes. Is there a

متن کامل

Relationship between Electrically Evoked Compound Action Potential Thresholds and Auditory, Language, and Speech Progress after Cochlear Implant Surgery

Introduction: Electrically evoked compound action potential (ECAP) is an objective auditory response that can be used in the programing of cochlear implants. The aims of this study were to monitor ECAP thresholds and auditory, language and speech progress for 6 months after cochlear implant surgery and to evaluate any relationship between them.   Materials and Methods: Ten children with a m...

متن کامل

Evaluation of the Slope of Amplitude Growth Function Changes of the Electrically Evoked Action Potential in Three Months after Receiving the Device in Children with Cochlear Implant

Objective: In neural response telemetry, intracochlear electrodes stimulate the auditory nerve and record the neural responses. The electrical stimulation is sent to the auditory nerve by an electrode and the resulted response, called electrically evoked compound action potential, is recorded by an adjacent electrode. The most important clinical applications of this test are evaluation and moni...

متن کامل

Comparison of electrically evoked whole-nerve action potential and electrically evoked auditory brainstem response thresholds in nucleus CI24R cochlear implant recipients.

In this study, differences between electrically evoked whole-nerve action potential (EAP) and electrically evoked auditory brainstem response (EABR) measurements within Nucleus CI24R cochlear implant recipients were evaluated. Precurved modiolus-hugging internal electrode arrays, such as the CI24R, are designed to provide more direct stimulation of neural elements of the modiolus. If the electr...

متن کامل

Analysis of electrically evoked compound action potential of the auditory nerve in children with bilateral cochlear implants.

INTRODUCTION The cochlear implant device has the capacity to measure the electrically evoked compound action potential of the auditory nerve. The neural response telemetry is used in order to measure the electrically evoked compound action potential of the auditory nerve. OBJECTIVE To analyze the electrically evoked compound action potential, through the neural response telemetry, in children...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Audiology & neuro-otology

دوره 10 2  شماره 

صفحات  -

تاریخ انتشار 2005