Reliability of Triggered EMG for Prediction of Safety during Pedicle Screw Placement in Adolescent Idiopathic Scoliosis Surgery

نویسندگان

  • Woo-Kie Min
  • Hyun-Joo Lee
  • Won-Ju Jeong
  • Chang-Wug Oh
  • Jae-Sung Bae
  • Hwan-Seong Cho
  • In-Ho Jeon
  • Chang-Hyun Cho
  • Byung-Chul Park
چکیده

STUDY DESIGN We performed a prospective study to evaluate the reliability of using triggered electromyography (EMG) for predicting pedicle wall breakthrough during the placement of pedicle screw in adolescent idiopathic scoliosis surgery. PURPOSE We wanted to correlate pedicle wall breakthrough with the triggered EMG threshold of stimulation and the postoperative computed tomography (CT) findings. OVERVIEW OF LITERATURE Pedicle wall breakthrough has been reported to be difficult to evaluate by radiographs. Triggered EMG had been found to be a more sensitive test to detect this breakthrough. METHODS Seven patients who underwent the insertion of 103 pedicle screws were evaluated. The triggered EMG activity was recorded from several muscles depending on the level of screw placement. The postoperative CT scans were read by a spine surgeon who was a senior fellow in orthopedics, and a musculoskeletal radiologist. RESULTS The mean age at surgery was 12.6 years (range, 11 to 17 years). The preoperative mean Cobb angle was 54.7° (range, 45 to 65°). There were 80 thoracic screws and 23 lumbar screws. All the screws had stimulation thresholds of ≥ 6 mA, except 3 screws with the stimulation threshold of < 6 mA. Ten screws (9.7%) showed violation of the pedicle wall on the postoperative CT scans. Five screws penetrated medially and another five penetrated laterally. No postoperative neurologic complications were noted in any of the seven patients. CONCLUSIONS Measuring the stimulation threshold of triggered EMG helps to assess the pedicle screw placement. Pedicle screws that had stimulation threshold of ≥ 6 mA were safe, with 90.3% reliability, as was assessed on the postoperative CT scans.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Accuracy of multilevel registration in image-guided pedicle screw insertion for adolescent idiopathic scoliosis.

STUDY DESIGN Retrospective clinical study. OBJECTIVE To assess the accuracy of multilevel registration for skip pedicle screw placement during image-guided, computer-assisted spine surgery, in the setting of adolescent idiopathic scoliosis (AIS). SUMMARY OF BACKGROUND DATA Computerized frameless stereotactic image-guidance has been used recently to improve pedicle screw placement accurately...

متن کامل

Safety of Pedicle Screws in Adolescent Idiopathic Scoliosis Surgery

To review existing publications on the safety of pedicle screw insertions in adolescent idiopathic scoliosis (AIS). Despite having increased risk for neurological and visceral injuries, the use of pedicle screws have led to increased correction rates in scoliosis surgery. A review was performed on topics pertinent to pedicle screw insertion in AIS, which included pedicle morphometry in AIS, str...

متن کامل

Pedicle Screw Instrumentation for Adolescent Idiopathic Scoliosis: The Insertion Technique, the Fusion Levels and Direct Vertebral Rotation

The pedicle is a power nucleus of the vertebra and offers a secure grip of all 3 columns. Pedicle screw instrumentation has advantages of rigid fixation with improved three-dimensional (3D) correction and it is accepted as a reliable method with a high margin of safety. Accurate placement of the pedicle screws is important to reduce possible irreversible complications. Many methods of screw ins...

متن کامل

Comparison of the pedicle screws placement between electronic conductivity device and normal pedicle finder in posterior surgery of scoliosis.

STUDY DESIGN Prospective randomized clinical trial. OBJECTIVE To compare the accuracy and time using of pedicle screw placement between electronic conductivity device (ECD) and normal pedicle finder (NPF) in posterior surgery of scoliosis, through a randomized clinical trial. SUMMARY OF BACKGROUND DATA Pedicle screw insertion for scoliosis correction can be associated with increased pedicle...

متن کامل

Evaluation of implant loosening following segmental pedicle screw fixation in adolescent idiopathic scoliosis: a 2 year follow-up with low-dose CT

BACKGROUND The long term radiological status of screw fixation following scoliosis surgery with all pedicle screw construct is not previously studied. AIM To evaluate the incidence of loosening (implant failure) evaluated with low-dose CT two years following scoliosis surgery. STUDY DESIGN Retrospective study. METHODS 81 consecutive patients with adolescent idiopathic scoliosis (AIS), age...