Diagnostic accuracy of serum ceruloplasmin in Wilson disease: determination of sensitivity and specificity by ROC curve analysis among ATP7B-genotyped subjects.

نویسندگان

  • Chloe M Mak
  • Ching-Wan Lam
  • Sidney Tam
چکیده

BACKGROUND A serum ceruloplasmin concentration below 0.20 g/L is conventionally considered as one of the major diagnostic criteria for Wilson disease. This decision threshold has not been fully validated for its diagnostic characteristics, however. In this study, we evaluated various decision thresholds of serum ceruloplasmin concentration in the diagnosis of Wilson disease based on genotype-verified Wilson disease patients, carriers, and normal individuals. METHODS Serum ceruloplasmin concentration was measured by a nephelometric method in 57 Wilson disease patients and 71 family members (49 heterozygotes and 22 wild-type homozygotes), a validation group of 25 subjects clinically suspected of Wilson disease, and 690 normal individuals. We performed ROC analysis using Analyze-it software and confirmed the genotypes by direct DNA sequencing of ATP7B. RESULTS Serum ceruloplasmin concentrations <0.20, 0.14, and 0.10 g/L showed positive predictive values of 48.3%, 100%, and 100%, respectively, and negative predictive values of 98.7%, 97.1%, and 91.9%. In the validation group, a serum ceruloplasmin threshold of 0.14 g/L rendered 100% sensitivity and specificity. Forty of 690 healthy subjects had serum ceruloplasmin concentrations <0.20 g/L; however, these 40 individuals had normal genotypes by DNA sequencing, and none of the 40 had ceruloplasmin concentrations <0.14 g/L. CONCLUSIONS The diagnostic accuracy for Wilson disease using a serum ceruloplasmin concentration of 0.14 g/L as the local decision threshold was better than that using a threshold of 0.20 g/L. We suggest that laboratories providing ceruloplasmin assays determine decision thresholds based on local populations.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Diagnostic Value of Ceruloplasmin in the Diagnosis of Pediatric Wilson's Disease

PURPOSE Measurement of serum ceruloplasmin level is the first step in screening for Wilson's disease (WD). Despite the rarity of WD in the general population, ceruloplasmin levels are routinely measured through hepatitis screening in both adults and children. Herein, we evaluated the diagnostic value of ceruloplasmin for the diagnosis of WD among children with hepatitis. METHODS We retrospect...

متن کامل

Truncating mutations in the Wilson disease gene ATP7B are associated with very low serum ceruloplasmin oxidase activity and an early onset of Wilson disease

BACKGROUND Mutations in the gene ATP7B cause Wilson disease, a copper storage disorder with a high phenotypic and genetic heterogeneity. We aimed to evaluate whether 'severe' protein-truncating ATP7B mutations (SMs) are associated with low serum ceruloplasmin oxidase activities and an early age of onset when compared to missense mutations (MMs). METHODS The clinical phenotype of 59 geneticall...

متن کامل

Application of adjusted-receiver operating characteristic curve analysis in combination of biomarkers for early detection of gestational diabetes mellitus

Introduction: In medical diagnostic field, evaluation of diagnostic accuracy of biomarkers or tests has always been a matter of concern. In some situations, one biomarker alone may not be sufficiently sensitive and specific for prediction of a disease. However, combining multiple biomarkers may lead to better diagnostic.  The aim of this study was to assess the performance of combination of bio...

متن کامل

ROC curve analysis: an example showing the relationships among serum lipid and apolipoprotein concentrations in identifying patients with coronary artery disease.

Clinical accuracy, defined as the ability to discriminate between states of health, is the fundamental property of any diagnostic test or system. It is readily expressed as clinical sensitivity and specificity, and elegantly represented by the receiver operating characteristic (ROC) curve. To demonstrate the use of ROC curves, we reexamine a study of the ability of serum lipid and apolipoprotei...

متن کامل

مقایسه مدل‌های رگرسیون لجستیک با تحلیل جداسازی در پیش‌بینی دیابت نوع 2

Background and Objectives: Diabetes is a chronic and common metabolic disease which has no curative treatment. Logistic regression (LR) is a statistical model for the analysis and prediction in multivariate statistical techniques. Discriminant analysis is a method for separating observations in terms of dependent variable levels which can allocate any new observation after making discriminating...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical chemistry

دوره 54 8  شماره 

صفحات  -

تاریخ انتشار 2008