The varying porosity of braided self-expanding stents and flow diverters: an experimental study.
نویسندگان
چکیده
BACKGROUND AND PURPOSE Braided self-expandable stents and flow diverters of uniform construction may develop zones of heterogeneous porosity in vivo. Unwanted stenoses may also occur at the extremities of the device. We studied these phenomena in dedicated benchtop experiments. MATERIALS AND METHODS Five braided devices of decreasing porosity were studied. To simulate discrepancies in diameters between the landing zones of the parent vessel and the aneurysm neck area, device extremities were inserted into silicone tubes of various diameters (2-3 mm), leaving the midportion free to react to experimental manipulations, which included axial approximation of the tubes (0-7 mm), and curvature (0-135°), with or without axial compression (0-2 mm). The length of the landing zone was sequentially decreased to study terminal device stenosis. RESULTS All devices adopted a conformation characterized by 3 different zones: bilateral landing zones, a middle compaction zone, and 2 transition zones. It is possible, during deployment, to compact stents and FDs to decrease porosity, but a limiting factor was the transition zone, which remained relatively unchanged and of higher porosity than the expansion zone. Length of the transition zone increased when devices were constrained in smaller tubes. Heterogeneities in porosity with compaction and curvatures were predictable and followed simple geometric rules. Extremity stenoses occurred increasingly with decreasing length of the landing zone. CONCLUSIONS Braided self-expandable devices show predictable changes in porosity according to device size, vessel diameter, and curvature. Adequate landing zones are required to prevent terminal device stenosis.
منابع مشابه
Flow diverters can occlude aneurysms and preserve arterial branches: a new experimental model.
BACKGROUND AND PURPOSE FDs are new intracranial stents designed to occlude aneurysms while preserving flow to jailed arterial branches. We tested this fundamental principle in a new aneurysm model. MATERIALS AND METHODS Canine lateral wall aneurysms, featuring a branch located immediately opposite the aneurysm, were created in 16 animals to study the effects on aneurysm or branch occlusion us...
متن کاملFigure 1: Schematic of the Idealized Sidewall Aneurysm Model (left) and Camera and Fluoroscopic Shots of the Five Devices Implanted in Silicone Models
Brain aneurysms are abnormal ballooning of intracranial arteries which when left untreated may result in fatal outcome for the patients. Neurovascular stents have been used extensively in stent-assisted coiling treatment of aneurysms. Treatment with flow diverters (low porosity, fine-mesh stents) is used for complicated aneurysm geometries not amenable to coiling. Angiographic analysis with mat...
متن کاملMechanical properties and in vitro degradation of bioabsorbable self-expanding braided stents.
The aim of this study was to characterize the mechanical and self-expansion properties of braided bioabsorbable stents. In total four different stents were manufactured from PLLA fibres using a braiding technique. The changes in radial pressure stiffness and diameter recovery of the stents were determined initially, and after insertion and release from a delivery device. The braided stents were...
متن کاملSelection of helical braided flow diverter stents based on hemodynamic performance and mechanical properties
BACKGROUND Although flow diversion is a promising procedure for the treatment of aneurysms, complications have been reported and it remains poorly understood. The occurrence of adverse outcomes is known to depend on both the mechanical properties and flow reduction effects of the flow diverter stent. OBJECTIVE To clarify the possibility of designing a flow diverter stent considering both hemo...
متن کاملParticle Imaging Velocimetry Evaluation of Intracranial Stents in Sidewall Aneurysm: Hemodynamic Transition Related to the Stent Design
We investigated the flow modifications induced by a large panel of commercial-off-the-shelf (COTS) intracranial stents in an idealized sidewall intracranial aneurysm (IA). Flow velocities in IA silicone model were assessed with and without stent implantation using particle imaging velocimetry (PIV). The use of the recently developed multi-time-lag method has allowed for uniform and precise meas...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- AJNR. American journal of neuroradiology
دوره 34 3 شماره
صفحات -
تاریخ انتشار 2013