Use of Cement Gun for Fixation of Tibia Component in Total Knee Arthroplasty
نویسندگان
چکیده
We evaluated the efficacy of a cement gun to improve the depth of cement penetration in total knee arthroplasty. Ninety-one consecutive patients from two hospitals were recruited for this study. For Group I cement was applied to the tibial baseplate and the proximal tibia with fingers. Group 2 had similar application of cement to the tibial baseplate but cement was pressurized into the proximal tibia using a cement gun.. The knee was kept extended until the cement hardened. Standard post-operative x-rays were reviewed to assess cement penetration into the proximal tibia. The mean cement penetration was 2.1 mm in Group 1 and 3.1 mm in Group 2 and the difference was statistically significant. The use of the cement gun improves cement penetration into the proximal tibia and facilitates early stability of the implant fixation to the bone.
منابع مشابه
How Much Bone Cement Is Utilized for Component Fixation in Primary Cemented Total Knee Arthroplasty?
Dear Editor, We read with great interest the timely study by Satish et al., How Much Bone Cement Is Utilized for Component Fixation in Primary Cemented Total Knee Arthroplasty? (1). We applaud the authors, yet we hope they can clarify some points to make this study more applicable to a wider population.
متن کاملHow Much Bone Cement Is Utilized for Component Fixation in Primary Cemented Total Knee Arthroplasty?
Background: No scientific evidence exists regarding the amount of bone cement used and discarded in primarycemented Total knee arthroplasty (TKA). The aim of this study was to identify the exact amount of bone cement utilizedfor component fixation in primary TKA.Methods: In a prospective study carried out at five centers, 133 primary cemented TKAs were performed. One packof 40g Palacos bone cem...
متن کاملEffect of hand packing versus cement gun pressurization on cement mantle in total knee arthroplasty.
BACKGROUND Gun pressurization in total knee arthroplasty (TKA) may result in better cement penetration than hand packing, leading to fewer tibial plate failures. We compared cement intrusion characteristics between vacuum mixing and gun pressurization versus hand mixing and packing in the proximal tibia among patients undergoing TKA. METHODS We analyzed 6-week radiographs from 77 consecutive ...
متن کاملIntrusion Characteristics of Three Bone Cements for Tibial Component of Total Knee Arthroplasty in a Cadaveric Bone Model.
The purpose of this study was to evaluate and compare the intrusion characteristics of Simplex-HV to Simplex-P and Palacos-R in cadaveric proximal tibial bone. Eighteen fresh-frozen cadaver proximal tibiae were examined with standard arthroplasty tibial cuts. Each tibia was randomly assigned to receive one of the three bone cements for use with finger packing technique. Sagittal sections were p...
متن کاملCement penetration using a tibial punch cement pressurizer in total knee arthroplasty.
This study describes a new tibial cement punch pressurizer to enhance cement penetration into the metaphyseal cancellous bone of the tibial plateau. Thirty knees of 15 patients undergoing simultaneous bilateral total knee arthroplasty (TKA) were randomly selected to receive the tibial cement punch pressurizer on one side, with the opposite side serving as control. Using computerized and radiogr...
متن کامل