Correction of Factitious Hyperkalemia in Hemolysed Specimens from Adult Emergency Department Using the Beckman Coulter Unicell DxC 880i® Analyzer Derived Factor

نویسنده

  • C. Domingues
چکیده

Hemolysis in ED (emergency department) patients is common due to difficult blood draws. Values of serum potassium (K) become falsely elevated secondary to release of intracellular contents. Objective: The aim of the study was to establish a correction factor for factitious elevated K in samples for de adult ED. Methods: We used samples from 125 adult ED patients, in which the 2nd sample was drawn due to hemolysis of the first tube. Results: Firstly, we derived a correction factor expressing an increase in potassium concentration in 0.21 mmol/L (95% confidence interval, 0.17-0.24 mmol/L with p < 0.01) for each hemolysis index increment. Conclusions: A reliable correction factor for factitious hyperkalemia in a clinical relevant range exists.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Evaluation of the Beckman Coulter UniCel DxH 800, Beckman Coulter LH 780, and Abbott Diagnostics Cell-Dyn Sapphire hematology analyzers on adult specimens in a tertiary care hospital.

We evaluated the new Beckman Coulter DxH 800 hematology analyzer (Beckman Coulter, Miami, FL) vs the Abbott Diagnostics Cell-Dyn Sapphire (Abbott Diagnostics, Santa Clara, CA) and Beckman Coulter LH 780 hematology analyzers using 430 adult specimens. The DxH 800 provided a CBC and differential that correlated well with those of the Sapphire and LH 780, with most parameters showing correlation c...

متن کامل

An 83-year-old woman with discordant urine protein results.

An 83-year-old woman presented to the emergency department with weakness and weight loss. A urinalysis was ordered. The urine was clear and light yellow. A urine protein estimate obtained with reagent strips (Clinitek Atlas ; Siemens) was 30 mg/dL; however, quantitative measurement of the total protein concentration (UniCel DxC 800; Beckman Coulter) of a second urine sample yielded a value of 9...

متن کامل

A lipemic sample?

CASE DESCRIPTION A 69-year-old male patient with nonischemic cardiomyopathy was admitted for biventricular systolic heart failure. Initial laboratory tests were cancelled due to a high lipemic index of 7 (Beckman UniCel® DxC). A redrawn sample yielded the same flag. The lipemic index did not change following ultracentrifugation in an airfuge. The sample was clear upon visual inspection but prec...

متن کامل

Hyperkalemia-induced complete heart block

Background: Potassium, as an extracellular ion, plays an important role in the electrophysiologic function of the myocardium and any change in extracellular concentration of this ion might have a marked impression upon myocyte electrophysiologic gain. High serum potassium levels are thought to impair pulse conduction in Purkinje fibers and ventricles more than that in the Atrio...

متن کامل

The effect of sodium bicarbonate and validation of beckman coulter AU680 analyzers for measuring total carbon dioxide (TCO2) concentrations in horse serum

This study evaluated the usage of Beckman Coulter AU680 analyzers for measurement of TCO 2 in horse serum, and the effect of sodium bicarbonate administrations on serum TCO 2 levels in resting horses. Treatment of horses with sodium bicarbonate did not result in any adverse events. Mean TCO 2 concentration was significantly higher from 1 to 8 h in the sodium bicarbonate-treated horses compared ...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2017