Minute Volume Viscometer
نویسندگان
چکیده
In the field of pharmaceutical research, injectable treatment therapy is currently one of the most highly examined methods of drug delivery. The continuation of this research is highly dependent on knowledge of fluid properties, specifically loss and storage modulus. Most drugs in the developmental stages of this method are extremely expensive and limited in quantity. Currently, devices that measure fluid properties require samples sizes larger than 150 μL. A device that would minimize the sample volume would be quite valuable to pharmaceutical companies. We have designed a device that will be able to quickly and accurately measure the loss and storage modulus of sample volumes of less than 20 μL. Three designs were considered throughout the planning of this project. One utilized polarized lenses with an alternating current to calculate the relative “brightness” of light through a sample liquid. A second method employed tuning forks vibrations as they propagate through a sample liquid to calculate waveform impedance. We have concluded that the optimal design for this project is one which calculates impedance of shear waves based on a loaded piezoelectric crystal.
منابع مشابه
Fabrication and Testing of Viscosity Measuring Instrument (Viscometer)
This paper presents the fabrication and testing of a simple and portable viscometer for the measurement of bulk viscosity of different Newtonian fluids. It is aimed at making available the instrument in local markets and consequently reducing or eliminating the prohibitive cost of importation. The method employed is the use of a D.C motor to rotate a disc having holes for infra-red light to pas...
متن کاملA New Viscometer
A viscometer is described, particularly adapted to the measurement of colloidal solutions and physiological liquids (serum, blood), which, although doing away with the capillary tube (which is difficult to handle and to clean), requires less than 1 cc. of liquid. The readings are rapid, from 30 seconds to 1 minute. The temperature is easily controlled and varied. The range is considerable and m...
متن کاملThe criteria for evaluating shear stress measuring range and the lowest measurable shear stress of rotational viscometer.
In the Guidelines for measurement of blood viscosity and erythrocyte deformability by ICSH in 1986, it points out that the viscometer for hemorheology should ideally be able to operate over a wide range of shear conditions, and for a constant shear-rate viscometer a high-shear measurement at 200/second and a low-shear measurement at 1/second (or shear-rates approximating to these values if they...
متن کاملEffects of red cell transfusion on cardiac output and blood flow velocities in cerebral and gastrointestinal arteries in premature infants.
Anaemia may increase the risk of tissue hypoxia in preterm infants. The effect of transfusion on circulation was studied in 33 preterm infants with a mean (SD) gestational age of 29 (5) weeks (range 26-34), birth weight 1153 (390) g (range 520-1840), and postnatal age of 48 (21) days (range 19-100). Packed cell volume, blood viscosity (capillary viscometer), cardiac output, and cerebral blood f...