Improved Detection of Sentinel Lymph Nodes in SPECT/CT Images Acquired Using a Low- to Medium-Energy General-Purpose Collimator.

نویسنده

  • Hiroto Yoneyama
چکیده

PURPOSE The use of the low-energy high-resolution (LEHR) collimator for lymphoscintigraphy causes the appearance of star-shaped artifacts at injection sites. The aim of this study was to confirm whether the lower resolution of the low- to medium-energy general-purpose (LMEGP) collimator is compensated by decrease in the degree of septal penetration and the reduction in star-shaped artifacts. METHODS A total of 106 female patients with breast cancer, diagnosed by biopsy, were enrolled in this study. Tc phytate (37 MBq, 1 mCi) was injected around the tumor, and planar and SPECT/CT images were obtained after 3 to 4 hours. When sentinel lymph nodes (SLNs) could not be identified from planar and SPECT/CT images by using the LEHR collimator, we repeated the study with the LMEGP collimator. RESULTS Planar imaging performed using the LEHR and LEHR + LMEGP collimators positively identified SLNs in 96.2% (102/106) and 99.1% (105/106) of the patients, respectively. Using combination of planar and SPECT/CT imaging with the LEHR and LEHR + LMEGP collimators, SLNs were positively identified in 97.2% (103/106) and 100% (106/106) of the patients, respectively. CONCLUSIONS The LMEGP collimator provided better results than the LEHR collimator because of the lower degree of septal penetration. The use of the LMEGP collimator improved SLN detection.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Usefulness of medium-energy collimator for sentinel node lymphoscintigraphy imaging in breast cancer patients.

UNLABELLED This study was performed to investigate the usefulness of a general-purpose medium-energy (ME) collimator for the accurate localization of the sentinel lymph node (SLN) in breast cancer patients. METHODS We compared phantom images and lymphoscintigraphy images obtained under different conditions for a patient with breast cancer. Comparisons were performed between 2 cameras, between...

متن کامل

Evaluation of iterative reconstruction method and attenuation correction on brain dopamine transporter SPECT using anthropomorphic striatal phantom

Objective(s): The aim of this study was to determine the optimal reconstruction parameters for iterative reconstruction in different devices and collimators for dopamine transporter (DaT) single-photon emission computed tomography (SPECT). The results were compared between filtered back projection (FBP) and different attenuation correction (AC) methods.Methods: An anthropomorphic striatal phant...

متن کامل

The Usefulness of SPECT/CT in Sentinel Node Mapping of Early Stage Breast Cancer Patients Showing Negative or Equivocal Findings on Planar Scintigraphy

Objective(s): This study sought to determine the diagnostic yield of SPECT/CT in localizing axillary sentinel lymph nodes (SLNs) in early breast cancer patients where planar scintigraphy (PS) was equivocal or negative.Methods: Prospective analysis of early stage breast cancer patients with nonpalpable axillary nodes undergoing SLN localization prior to nodal sampling for axillary staging. PS fi...

متن کامل

Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT Images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators

Objective(s): The purpose of this study was to examine the optimal reconstruction parameters for brain dopamine transporter SPECT images obtained with a fan beam collimator and compare the results with those obtained by using parallel-hole collimators.Methods: Data acquisition was performed using two SPECT/CT devices, namely a Symbia T6 and an Infinia Hawkeye 4 (device A and B) equipped with fa...

متن کامل

An anthropomorphic phantom study of brain dopamine transporter SPECT images obtained using different SPECT/CT devices and collimators.

UNLABELLED The aim of this study was to evaluate differences in dopamine transporter SPECT images among different SPECT/CT devices and to determine the most appropriate region of interest (ROI) for semiquantitative evaluation. METHODS An anthropomorphic striatal phantom was filled with (123)I solutions of different striatum-to-background radioactivity ratios. Data were acquired using 2 SPECT/...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Clinical nuclear medicine

دوره 40 7  شماره 

صفحات  -

تاریخ انتشار 2013