Phase transformation of a zirconia ceramic head after total hip arthroplasty.
نویسندگان
چکیده
We report two cases of surface deterioration of a zirconia ceramic femoral head associated with phase transformation after total hip arthroplasty. One head was retrieved at revision due to recurrent dislocation after six years and the other because of failure of the locking mechanism of the polyethylene liner after three years. The monoclinic content of the zirconia ceramics rose from 1% to about 30% on the surface of the heads. SEM revealed numerous craters indicating extraction of the zirconia ceramics at the surface. Surface roughness increased from an initial value of 0.006 microm up to 0.12 microm. This is the first report to show that phase transformation of zirconia ceramics causes deterioration of the surface roughness of the head in vivo after total hip arthroplasty.
منابع مشابه
Tribological and material analyses of retrieved alumina and zirconia ceramic heads correlated with polyethylene wear after total hip replacement.
It has been suggested that the wear of ultra-high molecular weight polyethylene (UHMWPE) in total hip replacement is substantially reduced when the femoral head is ceramic rather than metal. However, studies of alumina and zirconia ceramic femoral heads on the penetration of an UHMWPE liner in vivo have given conflicting results. The purpose of this study was to examine the surface characterist...
متن کاملFatal Cobalt Cardiomyopathy Following Revision Total Hip Arthroplasty – A Brief Report with Review of Literature
Ceramic bearing surfaces are being increasingly used in young patients undergoing total hip arthroplasty. However,failures have been reported including fractures even with the newer third generation ceramics. The recommendedtreatment for fracture of ceramic bearing surfaces is complete synovectomy and revision total hip arthroplasty.However, disappointing results have also bee...
متن کاملAlternative bearing surfaces in total hip arthroplasty: zirconia-alumina pairing. Contribution or caveat?
The aim of the present study was to evaluate the zirconia and alumina articulation in total hip arthroplasty in regard to clinical and radiological outcome. This is the first report concerning the clinical application of a hybrid ceramic articulation. Owing to ethical reasons, a limited number of patients was enrolled in the study. Ten consecutive patients with degenerative arthritis were rando...
متن کاملFracture of a Y-TZP ceramic femoral head. Analysis of a fault.
We studied factors contributing to the initiation of fracture and failure of a zirconia ceramic femoral head. The materials retrieved during a revision total hip replacement were submitted to either visual, stereomicroscopic and scanning electron microscopy (SEM) or SEM and energy-dispersive X-ray analysis. X-ray diffraction was performed in order to investigate the extent of tetragonal to mono...
متن کاملWear of a composite ceramic head caused by liner fracture.
Third-generation composite ceramics (eg, Delta; DePuy Orthopaedics, Warsaw, Indiana; Ceramtec, Plochingen, Germany) have greatly improved material characteristics compared with second-generation products. This case report presents a patient after total hip arthroplasty with a fractured ceramic liner and a heavily worn ceramic head (both third-generation ceramics) retrieved 9 months after surger...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
عنوان ژورنال:
- The Journal of bone and joint surgery. British volume
دوره 83 7 شماره
صفحات -
تاریخ انتشار 2001