Intra-Pelvic Migration of Sliding Hip Screw During Osteosynthesis of Hip Fracture: A Rare Avoidable Intraoperative Complication
نویسندگان
چکیده
INTRODUCTION Hip fractures, which are common among old patients, are classified into two groups: intracapsular and extracapsular fractures. Extracapsular fractures can be treated with extramedullary implants [e.g. dynamic hip screw (DHS)] or intramedullary nails. Dynamic hip screw is the treatment of choice in stable pertrochanteric fractures. Intrapelvic migration of the sliding screw is a very rare complication. CASE REPORT We report a case of a 90-year old Caucasian woman who had an unusual intraoperative complication during osteosynthesis procedure for extracapsular hip fracture fixation. In fact, the sliding hip screw went deep into the pelvis during surgery. This mishap required an abdominal surgical approach by the general surgeon to remove the screw. CONCLUSION Taking into consideration the poor quality of the bone in very old patients, we emphasize the importance of following every single step of the surgical procedure, in order to minimize the risk of this complication.
منابع مشابه
Intra-operative migration of dynamic hip screw into the pelvis.
The authors report a rare per-operative complication of intra-pelvic migration of dynamic hip screw, during osteosynthesis of an inter-trochanteric fracture. Possible reasons of migration are analyzed along with the importance of careful execution of the surgical technique to avoid such iatrogenic complications and medico-legal implications.
متن کاملHip screw lateral migration with no cut-out or non-union implication: a case report
Hip screw migration of peritrochanteric fracture fixation devices is a described complication in English literature. Medial migration occupies the majority of these cases whereas lateral migration is rare. We report the case of an 85-year-old woman whose intramedullary osteosynthesis of a trochanteric fracture was complicated by hip screw lateral migration. Mobilization was not influenced and n...
متن کاملIntrapelvic Protrusion of a Broken Guide Wire Fragment during Fixation of a Femoral Neck Fracture
Breakage of DHS guide wire during surgery and its migration into the pelvis through the hip joint is a rare complication and its removal can be very challenging for the surgeon. We share our experience of a similar case wherein we used an ‘iliofemoral’ approach to successfully remove the broken transfixing guidewire from the hip joint. Although iliofemoral approach is similar to the lateral win...
متن کاملMedial pelvic migration of the lag screw in a short gamma nail after hip fracture fixation: a case report and review of the literature
Hip fractures are a common injury among the elderly. Internal fixation with an intramedullary (IM) system has gained popularity for the treatment of intertrochanteric femur fractures. Multiple complications associated with IM fracture fixation have been described, however, we report a rare complication of medial pelvic migration of the lag screw of a short IM nail in a stable construct ten week...
متن کاملTRIGEN INTERTAN intramedullary nail versus sliding hip screw: a prospective, randomized multicenter study on pain, function, and complications in 684 patients with an intertrochanteric or subtrochanteric fracture and one year of follow-up.
BACKGROUND Both intramedullary nails and sliding hip screws are used with good results in the treatment of intertrochanteric and subtrochanteric fractures. The aim of our study was to assess whether use of the TRIGEN INTERTAN nail, as compared with a sliding hip screw, resulted in less postoperative pain, improved functional mobility, and reduced surgical complication rates for patients with an...