Monte Carlo investigation of electron beam output factors versus size of square cutout.
نویسندگان
چکیده
A major task in commissioning an electron accelerator is to measure relative output factors versus cutout size (i.e., cutout factors) for various electron beam energies and applicator sizes. We use the BEAM Monte Carlo code [Med Phys. 22, 503-524 (1995)] to stimulate clinical electron beams and to calculate the relative output factors for square cutouts. Calculations are performed for a Siemens MD2 linear accelerator with beam energies, 6, 9, 11, and 13 MeV. The calculated cutout factors for square cutouts in 10 X 10 cm2, 15 X 15 cm2, and 20 X 20 cm2 applicators at SSDs of 100 and 115 cm agree with the measurements made using a silicon diode within about 1% except for the smallest cutouts at SSD= 115 cm where they agree within 0.015. The details of each component of the dose, such as the dose from particles scattered off the jaws and the applicator, the dose from contaminant photons, the dose from direct electrons, etc., are also analyzed. The calculations show that inphantom side-scatter equilibrium is a major factor for the contribution from the direct component which usually dominates the output of a beam. It takes about 6 h of CPU time on a Pentium Pro 200 MHz computer to simulate an accelerator and additional 2 h to calculate the relative output factor for each cutout with a statistical uncertainty of 1%.
منابع مشابه
Dosimetry for Small, Irregular and Rectangular Field Size for Electron Beam Therapy
Purpose: This study was to investigate the effect of irregularity in shape of field size of electron beam on the percentage depth dose and the output factor. Study the variety of the Percentage Depth Dose (PDD) and the output factors for rectangular and small field sizes when one of field dimensions less than the Practical Range (Rp) of the electron energy used. Compare the measured output fact...
متن کاملMonte carlo simulation of varian clinac iX 10 MV photon beam for small field dosimetry
Background: The lack of lateral electronic disequilibrium (LED) becomes a main problem in small field. This factor affects the dose in target volume cannot predict correctly. In addition, utilization of high-energy linear accelerator (10 MV) can emit some unwanted particles (electron contamination). Therefore, the aim of this study was to characterize head linear accelerator (linac) Varian Clin...
متن کاملEvaluation of an electron Monte Carlo dose calculation algorithm for electron beams
The electron Monte Carlo (eMC) dose calculation algorithm of the Eclipse treatment planning system is based heavily upon Monte Carlo simulation of the linac head and modeling of the linac beam characteristics with minimal measurement of beam data. Commissioning of the eMC algorithm on multiple identical linacs provided a unique opportunity to systematically evaluate the algorithm with actual me...
متن کاملMeasurements of Photon Beam Flattening Filter Using an Anisotropic Analytical Algorithm and Electron Beam Employing Electron Monte Carlo
Introduction: This study aimed to report the measurement of photon and electron beams to configure the Analytical Anisotropic Algorithm and Electron Monte Carlo used in clinical treatment. Material and Methods: All measurements were performed in a large water phantom using a 3-dimensional scanning system (PTW, Germany). ...
متن کاملPhysical characteristics of electron beam from conventional and beam shaper IOERT applicator: A comparison study
Introduction: Intraoperative electron radiation therapy (IOERT) is one of the cancer treatment techniques that delivers high doses to tumor bed during surgery. IOERT can be performed by either conventional LINACs or dedicated IORT accelerators such as LIAC (Light Intraoperative Accelerator). Two types of applicators can be used with LIAC dedicated accelerator including conventi...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Medical physics
دوره 26 5 شماره
صفحات -
تاریخ انتشار 1999