Analytic expressions for the ultimate intrinsic signal-to-noise ratio and ultimate intrinsic specific absorption rate in MRI.

نویسندگان

  • E Kopanoglu
  • V B Erturk
  • E Atalar
چکیده

The ultimate intrinsic signal-to-noise ratio is the highest possible signal-to-noise ratio, and the ultimate intrinsic specific absorption rate provides the lowest limit of the specific absorption rate for a given flip angle distribution. Analytic expressions for ultimate intrinsic signal-to-noise ratio and ultimate intrinsic specific absorption rate are obtained for arbitrary sample geometries. These expressions are valid when the distance between the point of interest and the sample surface is smaller than the wavelength, and the sample is homogeneous. The dependence on the sample permittivity, conductivity, temperature, size, and the static magnetic field strength is given in analytic form, which enables the easy evaluation of the change in signal-to-noise ratio and specific absorption rate when the sample is scaled in size or when any of its geometrical or electrical parameters is altered. Furthermore, it is shown that signal-to-noise ratio and specific absorption rate are independent of the permeability of the sample. As a practical case and a solution example, a uniform, circular cylindrically shaped sample is studied.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Novel Techniques regarding Specific Absorption Rate and Field of View Reduction in Magnetic Resonance Imaging

NOVEL TECHNIQUES REGARDING SPECIFIC ABSORPTION RATE AND FIELD OF VIEW REDUCTION IN MAGNETIC RESONANCE IMAGING Emre Kopanoglu Ph.D. in Electrical and Electronics Engineering Supervisors: Prof. Dr. Ergin Atalar Assoc. Prof. Dr. Vakur B. Ertürk September, 2012 In this dissertation, novel strategies regarding the reduction of the specific absorption rate (SAR) and the reduction of the total scan ti...

متن کامل

Ultimate intrinsic signal-to-noise ratio in MRI.

A method to calculate the ultimate intrinsic signal-to-noise ratio (SNR) in a magnetic resonance experiment for a point inside an arbitrarily shaped object is presented. The ultimate intrinsic SNR is determined by body noise. A solution is obtained by optimizing the electromagnetic field to minimize total power deposition while maintaining a constant right-hand circularly polarized component of...

متن کامل

Evaluation of internal MRI coils using ultimate intrinsic SNR.

The upper bounds of the signal-to-noise ratio (also known as the "ultimate intrinsic signal-to-noise ratio" (UISNR)) for internal and external coils were calculated. In the calculation, the body was modeled as a dielectric cylinder with a small coaxial cylindrical cavity in which internal coils could be placed. The calculated UISNR values can be used as reference solutions to evaluate the perfo...

متن کامل

Digital Binary Phase-shift Keyed Signal Detector

We have developed the effective algorithm for detecting digital binary phase-shift keyed signals. This algorithm requires a small number of arithmetic operations over the signal period. It can be relatively easy implemented based on the modern programmable logic devices. It also provides high interference immunity by identifying signal presence when signal-to-noise ratio is much less that its w...

متن کامل

A Closed-Form Expression for Ultimate Intrinsic Signal-to-Noise Ratio in MRI

INTRODUCTION In magnetic resonance imaging (MRI), an increase in signal to noise ratio (SNR) provides higher quality images, making diagnosis easier, thus it should be maximized. The main noise sources in an MRI experiment are classified as preamplifier, coil and body noises. Preamplifier noise is small when ultra-low-noise amplifiers (ULNA) are utilized. Coil noise is also small in most applic...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Magnetic resonance in medicine

دوره 66 3  شماره 

صفحات  -

تاریخ انتشار 2011