A small and robust active beamstop for scattering experiments on high-brilliance undulator beamlines
نویسندگان
چکیده
A small active in-vacuum beamstop has been developed to monitor the flux of intense third-generation synchrotron X-ray beams protecting the downstream detector from the direct beam. Standard active beamstops, where a built-in diode directly absorbs the beam, have limitations in size and lifetime. In the present design, a silicon PIN diode detects the photons back-scattered from a cavity in the beamstop. This approach drastically reduces the radiation dose on the diode and thus increases its lifetime. The beamstop with a diameter of 2 mm has been fabricated to meet the requirements for the P12 bioSAXS beamline of EMBL Hamburg at PETRA III (DESY). The beamstop is in regular user operation at the beamline and displays a good response over the range of energies tested (6-20 keV). Further miniaturization of the diode is easily possible as its size is not limited by the PIN diode used.
منابع مشابه
Versatile sample environments and automation for biological solution X-ray scattering experiments at the P12 beamline (PETRA III, DESY)
A high-brilliance synchrotron P12 beamline of the EMBL located at the PETRA III storage ring (DESY, Hamburg) is dedicated to biological small-angle X-ray scattering (SAXS) and has been designed and optimized for scattering experiments on macromolecular solutions. Scatterless slits reduce the parasitic scattering, a custom-designed miniature active beamstop ensures accurate data normalization an...
متن کاملNumber of photons and brilliance of the radiation from a crystalline undulator
The scheme for accurate quantitative treatment of the radiation from a crystalline undulator in presence of the dechanneling and the photon attenuation is presented. The number of emitted photons and the brilliance of electromagnetic radiation generated by ultra-relativistic positrons channeling in a crystalline undulator are calculated for various crystals, positron energies and different bend...
متن کاملEstimation of Synchrotron Radiation Dose Outside The Hutch of SPring-8 Beamline
INTRODUCTION Synchrotron radiation facilities are to get high brilliance and high power density of synchrotron radiation. Especially, the third generation synchrotron radiation facilities such as the European Synchrotron Radiation Facility (ESRF)(1), The Advanced Photon Source (APS)(2), and SPring-8(3) can get extremely high brilliance, high power density, and high energy photons. The shields o...
متن کاملThe advanced photon source: performance and results from early operation.
The Advanced Photon Source at Argonne National Laboratory is now providing researchers with high-brilliance undulator radiation from below 1 keV to beyond 100 keV. All technical facilities and components are operational and have met design specifications. Fourteen research teams, with responsibility for 40 beamlines on the APS experiment hall floor, are currently installing beamline instrumenta...
متن کاملA short-pulse X-ray beamline for spectroscopy and scattering.
Experimental facilities for picosecond X-ray spectroscopy and scattering based on RF deflection of stored electron beams face a series of optical design challenges. Beamlines designed around such a source enable time-resolved diffraction, spectroscopy and imaging studies in chemical, condensed matter and nanoscale materials science using few-picosecond-duration pulses possessing the stability, ...