Influence of moisture accumulation in inline spacer on delivery of aerosol using metered-dose inhaler during mechanical ventilation.
نویسندگان
چکیده
BACKGROUND A practitioner questioned whether moisture that collected in the ventilator circuit and spacer affected the delivery of aerosol from a pressurized metered-dose inhaler (pMDI). An in vitro model was used to quantify the impact of accumulated humidity in a pMDI spacer and ventilator over time. METHODS A ventilator with an adult heated-wire ventilator circuit and humidifier was set to deliver adult settings. An impactor was placed between the endotracheal tube and the test lung to determine drug mass and mass median aerodynamic diameter of the aerosol delivered. An AeroVent pMDI spacer was placed in the inspiratory limb of the ventilator circuit and left in an open position. Eight actuations of HFA albuterol pMDI (720 microg) was administered at 1, 2, and 3 hours after the heater had reached equilibrium at 37 degrees C, and <10 min after turning off the heater/humidifier. The spacer was dried and returned to the heated circuit for additional testing. Samples were analyzed via spectrophotometer. One-way analysis of variance was applied (P<.05). RESULTS The delivered drug as a percent of emitted dose (mean+/-SD) was greater at hour one (23+/-2.1%) and with the dry spacer (21.8+/-3.3%) than at hours 2 and 3 or with humidifier off (11.4+/-3.8%, 12.3+/-0.8%, and 12.7+/-0.3%, respectively, P=.002). Mass median aerodynamic diameters with each comparison did not vary between conditions. Delivery efficiency was similar for the dry spacer and the spacer in the humidified circuit for one hour. However, once visible condensate occurred, drug delivery efficiency decreased by approximately 50%. CONCLUSIONS Aerosol delivery from a pMDI with spacer during mechanical ventilation was greater with a dry spacer and unchanged for the first hour after initiating heated humidification. Turning off the heated humidifier did not increase drug delivered.
منابع مشابه
Inhalation therapy with metered-dose inhalers and dry powder inhalers in mechanically ventilated patients.
Pressurized metered-dose inhalers (pMDIs) are commonly employed for administering bronchodilator aerosols to mechanically ventilated patients. Although it is feasible to employ dry powder inhalers in ventilator circuits, the presence of humidity in the ventilator circuit could reduce their efficiency. A complex array of factors influence drug delivery from pMDIs during mechanical ventilation, a...
متن کاملAerosol delivery during mechanical ventilation: from basic techniques to new devices.
Pressurized metered-dose inhalers (pMDIs) and nebulizers are routinely employed for aerosol delivery in mechanically ventilated patients. A significant proportion of the aerosol deposits in the ventilator circuit and artificial airway, thereby reducing the inhaled drug mass. Factors influencing aerosol delivery during mechanical ventilation differ from those in spontaneously breathing patients....
متن کاملBasic techniques for aerosol delivery during mechanical ventilation.
Among the devices employed for aerosol generation (metered-dose inhalers, nebulizers, and dry powder inhalers) only metered-dose inhalers and nebulizers are routinely employed for aerosol delivery to mechanically ventilated patients. The ventilator circuit and artificial airway were previously thought to be major barriers to effective aerosol delivery to mechanically ventilated patients. In the...
متن کاملBronchodilator delivery by metered-dose inhaler in mechanically ventilated COPD patients: influence of end-inspiratory pause.
The delivery of bronchodilators with a metered-dose inhaler (MDI) and a spacer in mechanically ventilated patients has become widespread practice. However, the various ventilator settings that influence the efficacy of MDI are not well established. Application of an end-inspiratory pause (EIP) during drug delivery has been suggested as one of the factors that might increase the effectiveness of...
متن کاملReconciling in vitro and in vivo measurements of aerosol delivery from a metered-dose inhaler during mechanical ventilation and defining efficiency-enhancing factors.
We attempted to resolve the discrepancies in reported data on aerosol deposition from a chlorofluorocarbon (CFC)-propelled metered-dose inhaler (MDI) during mechanical ventilation, obtained by in vivo and in vitro methodologies. Albuterol delivery to the lower respiratory tract was decreased in a humidified versus a dry circuit (16.2 versus 30.4%, respectively; p < 0.01). In 10 mechanically ven...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Respiratory care
دوره 54 10 شماره
صفحات -
تاریخ انتشار 2009