Baseline pressure errors (BPEs) extensively influence intracranial pressure scores: results of a prospective observational study
نویسندگان
چکیده
BACKGROUND Monitoring of intracranial pressure (ICP) is a cornerstone in the surveillance of neurosurgical patients. The ICP is measured against a baseline pressure (i.e. zero - or reference pressure). We have previously reported that baseline pressure errors (BPEs), manifested as spontaneous shift or drifts in baseline pressure, cause erroneous readings of mean ICP in individual patients. The objective of this study was to monitor the frequency and severity of BPEs. To this end, we performed a prospective, observational study monitoring the ICP from two separate ICP sensors (Sensors 1 and 2) placed in close proximity in the brain. We characterized BPEs as differences in mean ICP despite near to identical ICP waveform in Sensors 1 and 2. METHODS The study enrolled patients with aneurysmal subarachnoid hemorrhage in need of continuous ICP monitoring as part of their intensive care management. The two sensors were placed close to each other in the brain parenchyma via the same burr hole. The monitoring was performed as long as needed from a clinical perspective and the ICP recordings were stored digitally for analysis. For every patient the mean ICP as well as the various ICP wave parameters of the two sensors were compared. RESULTS Sixteen patients were monitored median 164 hours (ranges 70 - 364 hours). Major BPEs, as defined by marked differences in mean ICP despite similar ICP waveform, were seen in 9 of them (56%). The BPEs were of magnitudes that had the potential to alter patient management. CONCLUSIONS Baseline Pressure Errors (BPEs) occur in a significant number of patients undergoing continuous ICP monitoring and they may alter patient management. The current practice of measuring ICP against a baseline pressure does not comply with the concept of State of the Art. Monitoring of the ICP waves ought to become the new State of the Art as they are not influenced by BPEs.
منابع مشابه
The effect of baseline pressure errors on an intracranial pressure-derived index: results of a prospective observational study
BACKGROUND In order to characterize the intracranial pressure-volume reserve capacity, the correlation coefficient (R) between the ICP wave amplitude (A) and the mean ICP level (P), the RAP index, has been used to improve the diagnostic value of ICP monitoring. Baseline pressure errors (BPEs), caused by spontaneous shifts or drifts in baseline pressure, cause erroneous readings of mean ICP. Con...
متن کاملComparison of direct and indirect blood pressure in patients under esophagectomy surgery: a prospective observational study
Background: Arterial pressure is one of the most important physiological variables and often needs to be monitored repeatedly or continuously in perioperative period. Arterial pressure monitoring is one of the standard monitoring in operating room. During general anesthesia, blood pressure can be measured by using a noninvasive arterial pressure method or continuous invasive arterial pressure b...
متن کاملComparison of the Heated Humidified High-flow Nasal Cannula with Nasal Continuous Positive Airway Pressure as Primary Respiratory Support for Preterm Neonates: A Prospective Observational Study
Background: Heated humidified high-flow nasal cannula (HHHFNC) is gaining popularity in the management of respiratory distress in preterm neonates. However, it is not known whether it takes precedence over the gold standard nasal continuous positive airway pressure (NCPAP) in this age group as a primary mode of non-invasive ventilation (NIV). There is limited evidence ...
متن کاملA Colombian-based survey on ventriculostomy and intracranial pressure monitor placement practices
Objective: Monitoring of intracranial pressure (ICP) is a routine procedure in the neurosurgery field. Although the routine practice of monitoring ICP has decreased in recent years, the practice patterns have not been studied in Colombia. This study was designed to evaluate the current practice forventriculostomy and ICP monitoring by neurosurgeons in Colombia. Methods...
متن کاملCervical Interlaminar Epidural Injections at High Doses Do Not Increase Optic Nerve Sheath Diameter on a Long-Term Basis.
BACKGROUND Cervical epidural injection (CEI) is widely performed on patients with pain originating from the cervical spine. Studies have shown a good relationship between the optic nerve sheath diameter (ONSD) and the intracranial pressure (ICP). OBJECTIVE The aim of this study was to evaluate the changes in the ONSD as a non-invasive surrogate marker of ICP after CEI. STUDY DESIGN Prospect...