Comparison of intraocular pressure measurements using Goldmann tonometer, I-care pro, Tonopen XL, and Schiotz tonometer in patients after Descemet stripping endothelial keratoplasty
نویسندگان
چکیده
PURPOSE Post-Descemet stripping endothelial keratoplasty (DSEK) patients are prone for intraocular pressure (IOP) elevations and glaucoma. Corneal characteristics influence various IOP measuring devices in various ways. The aim of this study was to evaluate the agreement between four different IOP measuring devices: Goldmann applanation tonometer (GAT), I-care pro, Tonopen XL, and Schiotz tonometr in patients who underwent DSEK. METHODS This was a prospective comparative study using a convenience cohort of post-DSEK patients with compact grafts. Post-DSEK patients had IOP measured using GAT, I-care Pro, Tonopen XL, and Schiotz tonometer. Measurements were compared and agreement assessed. Wilcoxon signed-rank test was used for comparison of means as variables did not show a normal distribution. Bland-Altman plots were used for assessing agreement. RESULTS Thirty eyes of 24 patients were included in the study. Mean time from DSEK surgery was 25.31 ± 13.05 months. Mean IOP with GAT, I-care pro, Tonopen XL, and Schiotz tonometer was 13.99 ± 3.76, 13.92 ± 3.36, 13.31 ± 3.89, and 12.83 ± 4.07, respectively. GAT, I-care pro, and Tonopen XL had similar mean IOP measurements (P = 0.135 and P = 0.551, respectively), while Schiotz tonometry measurements were higher (P = 0.046). Bland-Altman plots show good agreement between GAT, Tonopen XL, and I-care pro. GAT and Schiotz tonometry show less agreement, with large variations in the differences of measured IOP. CONCLUSIONS IOP measurements in post-DSEK patients showed good agreement between GAT and either Tonopen XL or I-care pro. Schiotz tonometer has large variations in this patient group. IOP measurements and IOP difference between devices were not dependent on central corneal thickness.
منابع مشابه
Measurements of intraocular pressure by Goldmann tonometry, Tonopen XL, and the transpalpebral tonometer, TGDc-01, after penetrating keratoplasty: a comparativye study.
PURPOSE The aim of this study was to compare intraocular pressure (IOP) measurements obtained by Goldmann tonometry (GT), the Tonopen XL, and a new transpalpebral tonometer, TGDc-01, in eyes that had undergone penetrating keratoplasty (PKP). METHODS IOP was measured in post-PKP eyes by means of GT, Tonopen XL, and TGDc-01. Central corneal thickness measurements were also obtained for all eyes...
متن کاملSCIENTIFIC REPORT The effect of corneal thickness on intraocular pressure measurement in patients with corneal pathology
Background/aim: To compare intraocular pressure (IOP) measurements taken by the Goldmann applanation tonometer, the Tono-Pen and the ocular blood flow pneumotonometer in eyes with varying central corneal thickness (CCT) due to penetrating keratoplasty (PK), keratoconus (KC), and Fuchs’ endothelial dystrophy (FED). Methods: IOP was measured with the Goldmann applanation tonometer, Tono-Pen XL, a...
متن کاملThe effect of corneal thickness on intraocular pressure measurement in patients with corneal pathology.
BACKGROUND/AIM To compare intraocular pressure (IOP) measurements taken by the Goldmann applanation tonometer, the Tono-Pen and the ocular blood flow pneumotonometer in eyes with varying central corneal thickness (CCT) due to penetrating keratoplasty (PK), keratoconus (KC), and Fuchs' endothelial dystrophy (FED). METHODS IOP was measured with the Goldmann applanation tonometer, Tono-Pen XL, a...
متن کاملA Comparative Study of Intraocular Pressure Measurement Using Tonopen and Goldmann Applanation Tonometer
Background: Glaucoma is the world's leading cause of preventable irreversible blindness. The most important risk factor of glaucoma is increase in the intraocular pressure.Hence, timely screening by accurate IOP measurements is of utmost importance. In recent years, Tonopen has emerged as an alternative to Goldmann applanation tonometer which is the ‘gold standard’ for IOP measurement because o...
متن کاملTono-Pen tonometry in normal and in post-keratoplasty eyes.
Oculab Tono-Pen tonometry was compared with Goldmann applanation tonometry in 82 eyes of 82 patients with normal corneas and in 54 eyes of 54 patients who had undergone penetrating keratoplasty and whose corneas did not preclude the use of Goldmann tonometer. We found that the intraocular pressure (IOP) in 48% of the eyes with normal corneas and in 57% after keratoplasty has different measureme...
متن کامل