Mode of Onset of Spontaneous Ventricular Tachyarrhythmias from Implantable Cardioverter Defibrillator Electrograms
نویسندگان
چکیده
Aim: This work investigates the correlation between the cardiac electrical activity underlying spontaneous ventricular tachyarrhythmias and their modes of onset. Signals: We analyzed 81 episodes of spontaneous ventricular tachycardia (VT) and ventricular fibrillation (VF) lasting at least 5 seconds from 17 patients. The cardiologist classified VT/VF malignancy by considering fast VT as more malignant. We study 3 modes of onset: (i) premature ventricular contraction (PVC group); (ii) PVC preceded by a short-long-short cycle (SLS group); (iii) PVC preceded by a paced beat after a previous PVC compensatory pause (PM group). In 4 patients more than one mode of VT/VF onset is present, in the other 13 patients VT/VF always started with the same mode. Mathematical methods: Cardiac electrical activity is characterized by power spectral density and correlation dimension analyses; statistical analysis is carried on using one way ANOVA in order to test significant differences among the mean values of PVC, SLS and PM
منابع مشابه
Mode of onset of ventricular fibrillation in patients with Brugada syndrome detected by implantable cardioverter defibrillator therapy.
OBJECTIVES We sought to demonstrate the mode of spontaneous onset of ventricular fibrillation (VF) in patients with Brugada syndrome. BACKGROUND The electrophysiologic mechanisms of VF in Brugada syndrome have not been fully investigated. METHODS Nineteen patients (all male, mean age 47 +/- 12 years) with Brugada syndrome were treated with an implantable cardioverter defibrillator (ICD). Th...
متن کاملTachycardia induction due to inappropriate implantable cardioverter defibrillator therapy: what is the mechanism?
Antitachycardia therapies administered by implantable cardioverter defibrillators (ICDs) can occasionally prove proarrhythmic due to induction of ventricular tachyarrhythmias. In this report we present the case of a tachycardia induction as a result of inappropriate therapy delivery in an ICD recipient. Detailed analysis of the stored electrograms facilitated the delineation of the underlying m...
متن کاملArrhythmia/Electrophysiology Intracardiac Electrogram T-Wave Alternans/Variability Increases Before Spontaneous Ventricular Tachyarrhythmias in Implantable Cardioverter-Defibrillator Patients A Prospective, Multi-Center Study
Background—T-wave alternans (TWA) increases before ventricular tachycardia (VT) or fibrillation (VF), suggesting that it may warn of VT/VF in implantable cardioverter-defibrillator patients. Recently, we described a method for measuring alternans and nonalternans variability (TWA/V) from electrograms (EGMs) stored in implantable cardioverterdefibrillators before VT/VF. The goal of this prospect...
متن کاملImplantable cardioverter-defibrillator in a patient with dextrocardia situs inversus
Background: Dextrocardia is a congenital anomaly, which may have coexistent coronary artery disease (CAD), arrhythmias and conventional indications for device therapy. However, the implantation of transvenous leads can be technically challenging and the approach needs to be tailored to the patient's individual anatomy. Case presentation: A 54-year-old male with dextrocardia situs inversus and i...
متن کاملVentricular Tachycardia Recognized as Ventricular Fibrillation by an ICD Due to Double Counting of the QRS Complex
A case of ventricular tachycardia misdiagnosed as ventricular fibrillation due to double counting of the tachycardia’s wide QRS complex in an implantable cardioverter defibrillator (ICD) recipient is presented. As a result, an inappropriate shock was delivered by the device instead of anti-tachycardia pacing. Reprogramming of the detection criteria in a less sensitive manner resolved the proble...
متن کامل