Reliability of point-of-care testing of INR in acute stroke.

نویسندگان

  • Theresa L Green
  • Adnan Mansoor
  • Nancy Newcommon
  • Caroline Stephenson
  • Eileen Stewart
  • Michael D Hill
چکیده

BACKGROUND In the emergency department, portable point-of-care testing (POCT) coagulation devices may facilitate stroke patient care by providing rapid International Normalized Ratio (INR) measurement. The objective of this study was to evaluate the reliability, validity, and impact on clinical decision-making of a POCT device for INR testing in the setting of acute ischemic stroke (AIS). METHODS A total of 150 patients (50 healthy volunteers, 51 anticoagulated patients, 49 AIS patients) were assessed in a tertiary care facility. The INR's were measured using the Roche Coaguchek S and the standard laboratory technique. RESULTS The interclass correlation coefficient and 95% confidence interval between overall POCT device and standard laboratory value INRs was high (0.932 (0.69 - 0.78). In the AIS group alone, the correlation coefficient and 95% CI was also high 0.937 (0.59 - 0.74) and diagnostic accuracy of the POCT device was 94%. CONCLUSIONS When used by a trained health professional in the emergency department to assess INR in acute ischemic stroke patients, the CoaguChek S is reliable and provides rapid results. However, as concordance with laboratory INR values decreases with higher INR values, it is recommended that with CoaguChek S INRs in the > 1.5 range, a standard laboratory measurement be used to confirm the results.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Point-of-care international normalized ratio testing accelerates thrombolysis in patients with acute ischemic stroke using oral anticoagulants.

BACKGROUND AND PURPOSE Thrombolysis in patients using oral anticoagulants (OAC) and in patients for whom information on OAC status is not available is frequently delayed because the standard coagulation analysis procedure in central laboratories (CL) is time-consuming. By using point-of-care (POC) coagumeters, international normalized ratio (INR) values can be measured immediately at the bedsid...

متن کامل

Thrombolysis in stroke patients: Comparability of point-of-care versus central laboratory international normalized ratio

BACKGROUND In acute stroke patients, thrombolysis is one gold standard therapy option within the first four hours after the ischemic event. A contraindication for thrombolysis is an International Normalized Ratio (INR) value >1.7. Since time is brain, rapid and reliable INR results are fundamental. Aim was to compare INR values determined by central laboratory (CL) analyzer and Point-of-Care Te...

متن کامل

Clinical assessment of activities of daily living in acute stroke: Validation of the Persian version of Katz Index

Background: Katz Index is a well-known index for assessing basic activities of daily living. The aim was to determine validity and reliability of the Katz Index in Iranian patients with acute stroke. Methods: Eighty-seven patients (56 male, 31 female) with acute stroke (1-30 days post-stroke) participated in this psychometric properties study. Interval time for retest was 14 days. All partic...

متن کامل

Point-of-care test (POCT) INR: hope or illusion?

In the last decade, point-of-care tests were developed to provide rapid generation of test results. These tests have increasingly broad applications. In the area of hemostasis, the international normalized ratio, INR point-of-care test (POCT INR), is the main test of this new proposal. This test has great potential benefit in situations where the quick INR results influences clinical decision m...

متن کامل

Potential impact of point-of-care INR testing on intravenous thrombolysis.

Correspondence: Adriana Bastos Conforto; Rua Galeno de Almeida, 207 apto 23; 05410-030 São Paulo SP, Brasil; E-mail: [email protected] Conflict of intereste: There is no conflict of interest to declare. Received 01 June 2014 Accepted 09 June 2014 In this issue of Arquivos de Neuro-psiquiatria, Bruch et al. compared international normalized ratio (INR) measurements performed with point-...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The Canadian journal of neurological sciences. Le journal canadien des sciences neurologiques

دوره 35 3  شماره 

صفحات  -

تاریخ انتشار 2008