3D list-mode reconstruction for SPECT

نویسنده

  • L Bouwens
چکیده

| In previous work we developed a list-mode iterative reconstruction algorithm (LMIRA) for SPECT applied to two dimensional image reconstruction. In the 2D case we calculated the emission radiance distribution on the circular enclosure of the object. This part is the forward projection. In the backprojection we sampled the radiance on the intermediate layer using the geometric properties of the collimator. In 3D reconstruction we have to calculate the emission radiance distribution on the cylindrical enclosure. To characterize the radiance on the cylinder we need additional parameters to localize the position on the layer and the direction of the escaping photons, this results in a 4 dimensional function. The probabilities used in the LMIRA update equation are calculated from sampling the radiance with the cylindrical hole structure of the collimator. Keywords|List-mode, reconstruction, iterative.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Amplitude-based data selection for optimal retrospective reconstruction in micro-SPECT.

Respiratory motion can blur the tomographic reconstruction of positron emission tomography or single-photon emission computed tomography (SPECT) images, which subsequently impair quantitative measurements, e.g. in the upper abdomen area. Respiratory signal phase-based gated reconstruction addresses this problem, but deteriorates the signal-to-noise ratio (SNR) and other intensity-based quality ...

متن کامل

Influences of reconstruction and attenuation correction in brain SPECT images obtained by the hybrid SPECT/CT device: evaluation with a 3‐dimensional brain phantom

Objective: The aim of this study was to evaluate the influences of reconstruction and attenuation correction on the differences in the radioactivity distributions in 123I brain SPECT obtained by the hybrid SPECT/CT device. Methods: We used the 3-dimensional (3D) brain phantom, which imitates the precise structure of gray mater, white matter and bone regions. It was filled with 123I solution (20...

متن کامل

A New Approach for Quantitative Evaluation of Reconstruction Algorithms in SPECT

ABTRACT Background: In nuclear medicine, phantoms are mainly used to evaluate the overall performance of the imaging systems and practically there is no phantom exclusively designed for the evaluation of the software performance.  In this study the Hoffman brain phantom was used for quantitative evaluation of reconstruction techniques. The phantom is modified to acquire t...

متن کامل

List-mode SPECT reconstruction using empirical likelihood

. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Chapter

متن کامل

Evaluation of the Reconstruction Parameters of Brain Dopamine Transporter SPECT Images Obtained by a Fan Beam Collimator: A Comparison with Parallel-hole Collimators

Objective(s): The purpose of this study was to examine the optimal reconstruction parameters for brain dopamine transporter SPECT images obtained with a fan beam collimator and compare the results with those obtained by using parallel-hole collimators.Methods: Data acquisition was performed using two SPECT/CT devices, namely a Symbia T6 and an Infinia Hawkeye 4 (device A and B) equipped with fa...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2001