Banding and Step-Stair Artifacts on the Cardiac-CT Caused By Pseudo-Ectopic Beats.

نویسندگان

  • Amolak Singh
  • Yash Sethi
  • Sonya Watkins
  • Angela Youtsey
  • Angie Thomas
چکیده

Step-stair and banding artifacts may result from irregular ventricular rhythm caused by atrial fibrillation or premature ectopic ventricular contractions. In the case reported here, severe banding and misalignment artifacts occurred due to electrocardiographic noise mimicking ectopic beats. Severe EKG noise or pseudo-ectopic beats may cause rare but serious artifacts during cardiac-CT acquisition. Vendor-provided software for correcting ectopic beats can be used to remove pseudo-ectopic beats and eliminate artifacts caused by the EKG noise. All efforts should be made to prevent this kind of problem from happening in the first place.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Is correction for metallic artefacts mandatory in cardiac SPECT/CT imaging in the presence of pacemaker and implantable cardioverter defibrillator leads?

Introduction: Metallic artifacts due to pacemaker/ implantable cardioverter defibrillator (ICD) leads in CT images can produce artifactual uptake in cardiac SPECT/CT images. The aim of this study was to determine the influence of the metallic artifacts due to pacemaker and ICD leads on myocardial SPECT/CT imaging. Methods: The study included 9 patients wh...

متن کامل

Locations of ectopic beats coincide with spatial gradients of NADH in a regional model of low-flow reperfusion.

We studied the origins of ectopic beats during low-flow reperfusion after acute regional ischemia in excised rat hearts. The left anterior descending coronary artery was cannulated. Perfusate was delivered to the cannula using an high-performance liquid chromatography pump. This provided not only precise control of flow rate but also avoided mechanical artifacts associated with vessel occlusion...

متن کامل

Effects of Acetyl-L-Carnitine on Cardiac Arrhythmias and Infarct Size in Ischemic-Reperfused Isolated Rat Heart

Objective(s) This study aimed to examine whether acetyl-L-carnitine (ALC) was able to reduce cardiac arrhythmias and infarct size in the ischemic-reperfused isolated rat heart. Materials and Methods The isolated hearts were mounted on a Langendorff apparatus then perfused by a modified Krebs-Henseleit solution during 30 min regional ischemia and 120 min reperfusion (control) or by enriched K...

متن کامل

Sudden cardiac death: role of left ventricular dysfunction.

Early studies from the Coronary Drug Project of sudden cardiac death in patients surviving an episode of myocardial infarction called attention to the importance of ventricular ectopic beats.' In that study, complex ventricular ectopic beats recorded on a 12-hour Holter electrocardiogram were found to be associated with an excessive risk of sudden cardiac death. When the relationship was examin...

متن کامل

Artifact reduction techniques in Cone Beam Computed Tomography (CBCT) imaging modality

Introduction: Cone beam computed tomography (CBCT) was introduced and became more common based on its low cost, fast image procedure rate and low radiation dose compared to CT. This imaging modality improved diagnostic and treatment-planning procedures by providing three-dimensional information with greatly reduced level of radiation dose compared to 2D dental imaging modalitie...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of radiology case reports

دوره 3 6  شماره 

صفحات  -

تاریخ انتشار 2009