Safety Studies for a 44-Channel Suprachoroidal Retinal Prosthesis: A Chronic Passive Study.
نویسندگان
چکیده
Purpose Following successful clinical outcomes of the prototype suprachoroidal retinal prosthesis, Bionic Vision Australia has developed an upgraded 44-channel suprachoroidal retinal prosthesis to provide a wider field of view and more phosphenes. The aim was to evaluate the preclinical passive safety characteristics of the upgraded electrode array. Methods Ten normal-sighted felines were unilaterally implanted with an array containing platinum electrodes (44 stimulating and 2 returns) on a silicone carrier near the area centralis. Clinical assessments (color fundus photos, optical coherence tomography, full-field electroretinography, intraocular pressure) were performed under anesthesia prior to surgery, and longitudinally for up to 20 weeks. Histopathology grading of fibrosis and inflammation was performed in two animals at 13 to 15 weeks. Results Eight animals showed safe electrode array insertion (good retinal health) and good conformability of the array to the retinal curvature. Eight animals demonstrated good mechanical stability of the array with only minor (<2 disc diameters) lateral movement. Four cases of surgical or stability complications occurred due to (1) bulged choroid during surgery, (2) hemorrhage from a systemic bleeding disorder, (3) infection, and (4) partial erosion of thin posterior sclera. There was no change in retinal structure or function (other than that seen at surgery) at endpoint. Histopathology showed a mild foreign body response. Electrodes were intact on electrode array removal. Conclusions The 44-channel suprachoroidal electrode array has an acceptable passive safety profile to proceed to clinical trial. The safety profile is expected to improve in human studies, as the complications seen are specific to limitations (anatomic differences) with the feline model.
منابع مشابه
Chronic Electrical Stimulation with a Suprachoroidal Retinal Prosthesis: A Preclinical Safety and Efficacy Study
PURPOSE To assess the safety and efficacy of chronic electrical stimulation of the retina with a suprachoroidal visual prosthesis. METHODS Seven normally-sighted feline subjects were implanted for 96-143 days with a suprachoroidal electrode array and six were chronically stimulated for 70-105 days at levels that activated the visual cortex. Charge balanced, biphasic, current pulses were deliv...
متن کاملFirst-in-Human Trial of a Novel Suprachoroidal Retinal Prosthesis
UNLABELLED Retinal visual prostheses ("bionic eyes") have the potential to restore vision to blind or profoundly vision-impaired patients. The medical bionic technology used to design, manufacture and implant such prostheses is still in its relative infancy, with various technologies and surgical approaches being evaluated. We hypothesised that a suprachoroidal implant location (between the scl...
متن کاملFactors affecting perceptual thresholds in a suprachoroidal retinal prosthesis.
PURPOSE The suprachoroidal location for a retinal prosthesis provides advantages over other locations in terms of a simplified surgical procedure and a potentially more stable electrode-neural interface. The aim of this study was to assess the factors affecting perceptual thresholds, and to optimize stimulus parameters to achieve the lowest thresholds in patients implanted with a suprachoroidal...
متن کاملImplantation of multiple suprachoroidal electrode arrays in rabbits
Purpose Epiretinal and subretinal prosthesis have been shown to be a valid way to provide some vision to patients with advanced outer retinal degeneration and profound vision loss. However, the field of vision for these patients is markedly limited by the area occupied by the electrode array. In this study, we aimed to evaluate the feasibility of implantation of multiple suprachoroidal electrod...
متن کاملOne-Year Outcome of 49-Channel Suprachoroidal-Transretinal Stimulation Prosthesis in Patients With Advanced Retinitis Pigmentosa.
Purpose To determine the feasibility of a 49-channel suprachoroidal-transretinal stimulation (STS) retinal prosthesis that was implanted for 1 year. Methods Three patients with advanced retinitis pigmentosa (RP) and with less than or equal to hand motion (HM) vision were studied. A 49-channel STS system was implanted in a scleral pocket, and the stability of the prosthesis, adverse events, an...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Investigative ophthalmology & visual science
دوره 59 3 شماره
صفحات -
تاریخ انتشار 2018