Flat-panel digital mammography system: contrast-detail comparison between screen-film radiographs and hard-copy images.
نویسندگان
چکیده
PURPOSE To compare the contrast-detail (CD) characteristics of screen-film (SF) and postprocessed digital images by using a phantom-based method. MATERIALS AND METHODS Images of a CD phantom with polymerized methyl methacrylate were acquired with SF and full-field digital mammography systems at matched exposure conditions. A four-alternative forced-choice experiment was conducted with seven observers participating in the study. Each observer was required to identify randomly located disks in phantom images from which detection curves were computed. The CD diagrams for the SF and digital systems were estimated from the detection curves and compared at 50% and 62.5% threshold levels. Furthermore, a theoretic model was used to estimate the CD performance of the SF and digital systems. RESULTS Analysis of covariance for mixed models was used with the natural logarithm of disk thickness as the dependent variable, the natural logarithm of disk diameter as the covariate, and the observer as a random factor. The results of statistical analysis indicated significant differences between the CD characteristics of SF and digital mammographic images at both 50% (P <.001) and 62.5% (P <.001) detection thresholds. CONCLUSION The authors conclude that digital CD curves, on average, exhibit threshold contrast characteristics that are lower (better) than those of SF mammography.
منابع مشابه
A Comparison of the effectiveness of Mammographic Film-Screen and Standard Film-Screen in the Detection of Small Bone Fractures
The use of mammography film-screen is limited in general radiography. The purpose of this study was to compare the effectiveness of mammographic film-screen and standard film-screen systems in the detection of small bone fractures. Radiographs were taken from patients' extremities and neck areas using mammography film-screen and standard film-screen (n=57 each). Fourteen other radiographs were ...
متن کاملComparison of contrast detail curves of full field digital with screen film breast phantom images.
In this investigation, we imaged a standard breast phantom and compared the contrast detail curves from a prototype full breast digital mammography system with the corresponding curves for a conventional, analog screen-film system. The full breast digital system exhibited superior contrast detail detectability. The results from this study will be used to plan future clinical evaluations compari...
متن کاملPhysical and clinical comparison between a screen-film system and a dual-side reading mammography-dedicated computed radiography system.
BACKGROUND Digital mammography systems, thanks to a physical performance better than conventional screen-film units, have the potential of reducing the dose to patients, without decreasing the diagnostic accuracy. PURPOSE To achieve a physical and clinical comparison between two systems: a screen-film plate and a dual-side computed radiography system (CRM; FUJIFILM FCR 5000 MA). MATERIAL AN...
متن کاملComparison of calcification specificity in digital mammography using soft-copy display versus screen-film mammography.
OBJECTIVE The purpose of this study was to compare specificity in the interpretation of calcifications in soft-copy reviewing of digital mammograms versus hard-copy reviewing of screen-film mammograms. MATERIALS AND METHODS A total of 130 consecutive cases with calcifications (44 malignant and 86 benign) that had been evaluated with needle or surgical biopsy were collected. Both screen-film m...
متن کاملEpidemiology, Prevention, Screening
Background: Over 90% of mammography machines in Japan have already become digital, however, almost three-fourths of them utilize computed radiography (CR). The majority of them necessitate hard copy diagnosis. Therefore, reliability of soft-copy interpretation of CR mammography is still controversial. The purpose of this study is to assess the usefulness and problems of soft-copy interpretation...