Performance of the Vitek MS matrix-assisted laser desorption ionization time-of-flight mass spectrometry system for identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories.
نویسندگان
چکیده
We evaluated the performance of the Vitek MS for identification of Gram-positive cocci routinely isolated in clinical microbiology laboratories. With a total of 424 well-characterized isolates, the results of the Vitek MS were compared to those of conventional methods and 16S rRNA gene sequencing. The Vitek MS correctly identified 97.9 % of the isolates tested to species level. The Vitek MS correctly identified the species of 97.2 % of the staphylococci (95.9 % of coagulase-negative staphylococci), 97.8 % of the streptococci, and 100 % of the enterococci. For the identification of Gram-positive cocci isolates, the overall concordance rate between conventional identification and the Vitek MS was 94.5 %. The Vitek MS matrix-assisted laser desorption ionization time-of-flight mass spectrometry (MALDI-TOF MS) system can be a reliable and rapid method for the identification of most relevant Gram-positive cocci. In addition, expanding the database of the Vitek MS, especially for coagulase-negative staphylococci, is needed to enhance the performance of the Vitek MS.
منابع مشابه
Head-to-head comparison of Microflex LT and Vitek MS systems for routine identification of microorganisms by MALDI-TOF mass spectrometry in Chile
BACKGROUND Matrix-assisted laser desorption ionization-time of flight (MALDI-TOF) mass spectrometry (MS) is a new and revolutionary identification method for microorganisms and has recently been introduced into clinical microbiology in many industrialized countries in Europe and North America. OBJECTIVES Our study aimed to compare the performance and practicality of two commercial MALDI-TOF M...
متن کاملPerformance of the VITEK MS matrix-assisted laser desorption ionization-time of flight mass spectrometry system for rapid bacterial identification in two diagnostic centres in China.
Matrix-assisted laser desorption ionization-time of flight (MALDI-TOF) MS systems was not officially launched for diagnostic use in clinical microbiology laboratories in China until 2012. Here, we report the findings from the first large-scale evaluation study of VITEK MS for routine bacterial identification in two major diagnostic centres in Beijing and Hong Kong. A total of 2266 unique isolat...
متن کاملDirect Identification of Urinary Tract Pathogens From Urine Samples Using the Vitek MS System Based on Matrix-Assisted Laser Desorption Ionization-Time of Flight Mass Spectrometry
BACKGROUND We evaluated the coincidence rate between Vitek MS system (bioMérieux, France) and Vitek 2 in identifying uropathogens directly from urine specimens. METHODS Urine specimens submitted to our microbiology laboratory between July and September 2013 for Gram staining and bacterial culture were analyzed. Bacterial identification was performed by using the conventional method. Urine spe...
متن کاملMulticenter evaluation of the Vitek MS matrix-assisted laser desorption ionization-time of flight mass spectrometry system for identification of Gram-positive aerobic bacteria.
Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF) is gaining momentum as a tool for bacterial identification in the clinical microbiology laboratory. Compared with conventional methods, this technology can more readily and conveniently identify a wide range of organisms. Here, we report the findings from a multicenter study to evaluate the Vitek MS v2.0 sy...
متن کاملApplication of matrix-assisted laser desorption ionization-time of flight mass spectrometry for identification of the fastidious pediatric pathogens Aggregatibacter, Eikenella, Haemophilus, and Kingella.
The accuracy of matrix-assisted laser desorption-ionization time of flight mass spectrometry (MALDI-TOF MS) in the identification of Haemophilus, Aggregatibacter, Cardiobacterium, Eikenella, and Kingella (HACEK) species was compared to that of phenotypic methods (Remel RapID and Vitek 2). Overall, Vitek MS correctly identified more isolates, incorrectly identified fewer isolates, and failed to ...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of medical microbiology
دوره 62 Pt 9 شماره
صفحات -
تاریخ انتشار 2013