Quantifying CBF with pulsed ASL: technical and pulse sequence factors.
نویسنده
چکیده
We summarize here current methods for the quantification of CBF using pulsed arterial spin labeling (ASL) methods. Several technical issues related to CBF quantitation are described briefly, including transit delay, signal from larger arteries, radio frequency (RF) slice profiles, magnetization transfer, tagging efficiency, and tagging geometry. Many pulsed tagging schemes have been devised, which differ in the type of tag or control pulses, and which have various advantages and disadvantages for quantitation. Several other modifications are also available that can be implemented as modules in an ASL pulse sequence, such as varying the wash-in time to estimate the transit delay. Velocity-selective ASL (VS-ASL) uses a new type of pulse labeling in which inflowing arterial spins are tagged based on their velocity rather than their spatial location. In principle, this technique may allow ASL measurement of cerebral blood flow (CBF) that is insensitive to transit delays.
منابع مشابه
Quantifying CBF with arterial spin labeling.
The basic principles of measuring cerebral blood flow (CBF) using arterial spin labeling (ASL) are reviewed. The measurement is modeled by treating the ASL method as a magnetic resonance imaging (MRI) version of a microsphere study, rather than a diffusible tracer study. This approach, particularly when applied to pulsed ASL (PASL) experiments, clarifies that absolute calibration of CBF primari...
متن کاملTiming dependence of peripheral pulse-wave-triggered pulsed arterial spin labeling.
Arterial spin labeling (ASL) has been developed into a useful technique that is capable of quantifying noninvasively local cerebral blood flow (CBF) using the water molecules in arterial blood as diffusible tracers. Pulsed ASL (PASL) is more strongly affected than continuous ASL (CASL) by cardiac pulsation, because the tag bolus is shorter than the cardiac cycle in most cases. No reports have y...
متن کاملVelocity-selective arterial spin labeling.
In pathologies in which slow or collateral flow conditions may exist, conventional arterial spin labeling (ASL) methods that apply magnetic tags based on the location of arterial spins may not provide robust measures of cerebral blood flow (CBF), as the transit delay for the delivery of blood to target tissues may far exceed the relaxation time of the tag. Here we describe current methods for A...
متن کاملQuantifying Cerebellum Grey Matter and White Matter Perfusion Using Pulsed Arterial Spin Labeling
To facilitate quantification of cerebellum cerebral blood flow (CBF), studies were performed to systematically optimize arterial spin labeling (ASL) parameters for measuring cerebellum perfusion, segment cerebellum to obtain separate CBF values for grey matter (GM) and white matter (WM), and compare FAIR ASST to PICORE. Cerebellum GM and WM CBF were measured with optimized ASL parameters using ...
متن کاملArterial Transit Time Mapping Obtained by Pulsed Continuous 3D ASL Imaging with Multiple Post-Label Delay Acquisitions: Comparative Study with PET-CBF in Patients with Chronic Occlusive Cerebrovascular Disease
Arterial transit time (ATT) is most crucial for measuring absolute cerebral blood flow (CBF) by arterial spin labeling (ASL), a noninvasive magnetic resonance (MR) perfusion assessment technique, in patients with chronic occlusive cerebrovascular disease. We validated ASL-CBF and ASL-ATT maps calculated by pulsed continuous ASL (pCASL) with multiple post-label delay acquisitions in patients wit...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of magnetic resonance imaging : JMRI
دوره 22 6 شماره
صفحات -
تاریخ انتشار 2005