A Novel Approach for Removing the Hook Effect Artefact from Electrical Bioimpedance Spectroscopy Measurements

نویسندگان

  • R Buendia
  • F Seoane
  • R Gil-Pita
چکیده

Very often in Electrical Bioimpedance (EBI) spectroscopy measurements the presence of stray capacitances creates a measurement artefact commonly known as Hook Effect. Such an artefact creates a hook-alike deviation of the EBI data noticeable when representing the measurement on the impedance plane. Such Hook Effect is noticeable at high frequencies but it also causes a data deviation at lower measurement frequencies. In order to perform any accurate analysis of the EBI spectroscopy data, the influence of the Hook Effect must be removed. An established method to compensate the hook effect is the well known Td compensation, which consists on multiplying the obtained spectrum, Zmeas(ω) by a complex exponential in the form of exp[jωTd]. Such a method cannot correct entirely the Hook Effect since the hook-alike deviation occurs a broad frequency range in both magnitude and phase of the measured impedance, and by using a scalar value for Td. First a scalar only modifies the phase of the measured impedance and second, a single value can truly corrects the Hook Effect only at a single frequency. In addition, the process to select a value for the scalar Td by an iterative process with the aim to obtain the best Cole fitting lacks solid scientific grounds. In this work the Td compensation method is revisited and a modified approach for correcting the Hook Effect including a novel method for selecting the correcting values is proposed. The initial validation results confirm that the proposed method entirely corrects the Hook Effect at all frequencies.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Paper I Experimental validation of a method for removing the capacitive leakage artifact from electrical bioimpedance spectroscopy measurements

Often when performing electrical bioimpedance (EBI) spectroscopy measurements, the obtained EBI data present a hook-like deviation, which is most noticeable at high frequencies in the impedance plane. The deviation is due to a capacitive leakage effect caused by the presence of stray capacitances. In addition to the data deviation being remarkably noticeable at high frequencies in the phase and...

متن کامل

Experimental validation of a method for removing the capacitive leakage artifact

Often when performing electrical bioimpedance (EBI) spectroscopy measurements, the obtained EBI data present a hook-like deviation, which is most noticeable at high frequencies in the impedance plane. The deviation is due to a capacitive leakage effect caused by the presence of stray capacitances. In addition to the data deviation being remarkably noticeable at high frequencies in the phase and...

متن کامل

An analog front-end enables electrical impedance spectroscopy system on-chip for biomedical applications.

The increasing number of applications of electrical bioimpedance measurements in biomedical practice, together with continuous advances in textile technology, has encouraged several researchers to make the first attempts to develop portable, even wearable, electrical bioimpedance measurement systems. The main target of these systems is personal and home monitoring. Analog Devices has made avail...

متن کامل

Cole Parameter Estimation from the Modulus of the Electrical Bioimpeadance for Assessment of Body Composition. A Full Spectroscopy Approach

Activities around applications of Electrical Bioimpedance Spectroscopy (EBIS) have proliferated in the past decade significantly. Most of these activities have been focused in the analysis of the EBIS measurements, which eventually might enable novel applications. In Body Composition Assessment (BCA), the most common analysis approach currently used in EBIS is based on the Cole function, which ...

متن کامل

A Novel Approach for Estimation of Electrical Bioimpedance: Total Least Square

There are several methods used for ACimpedance estimation in Electrical Bioimpedance measurements. In this paper we propose a novel method for digital estimation of electrical impedance based in the Total Least Square (TLS) technique and we carry out a performance comparison between the proposed method and the typical Digital Sine Correlation (DSC) method. The TLS method has been implemented us...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2010