An analytic model for accurate spring constant calibration of rectangular atomic force microscope cantilevers
نویسندگان
چکیده
Spring constant calibration of the atomic force microscope (AFM) cantilever is of fundamental importance for quantifying the force between the AFM cantilever tip and the sample. The calibration within the framework of thin plate theory undoubtedly has a higher accuracy and broader scope than that within the well-established beam theory. However, thin plate theory-based accurate analytic determination of the constant has been perceived as an extremely difficult issue. In this paper, we implement the thin plate theory-based analytic modeling for the static behavior of rectangular AFM cantilevers, which reveals that the three-dimensional effect and Poisson effect play important roles in accurate determination of the spring constants. A quantitative scaling law is found that the normalized spring constant depends only on the Poisson's ratio, normalized dimension and normalized load coordinate. Both the literature and our refined finite element model validate the present results. The developed model is expected to serve as the benchmark for accurate calibration of rectangular AFM cantilevers.
منابع مشابه
Calibration of colloid probe cantilevers using the dynamic viscous response of a confined liquid
Articles you may be interested in Adhesive-free colloidal probes for nanoscale force measurements: Production and characterization Rev. Improved in situ spring constant calibration for colloidal probe atomic force microscopy Rev. Accurate noncontact calibration of colloidal probe sensitivities in atomic force microscopy Rev. A calibration method for lateral forces for use with colloidal probe f...
متن کاملMeasuring the spring constant of atomic force microscope cantilevers: thermal fluctuations and other methods
Knowledge of the interaction forces between surfaces gained using an atomic force microscope (AFM) is crucial in a variety of industrial and scientific applications and necessitates a precise knowledge of the cantilever spring constant. Many methods have been devised to experimentally determine the spring constants of AFM cantilevers. The thermal fluctuation method is elegant but requires a the...
متن کاملSpring constant calibration of atomic force microscope cantilevers of arbitrary shape.
The spring constant of an atomic force microscope cantilever is often needed for quantitative measurements. The calibration method of Sader et al. [Rev. Sci. Instrum. 70, 3967 (1999)] for a rectangular cantilever requires measurement of the resonant frequency and quality factor in fluid (typically air), and knowledge of its plan view dimensions. This intrinsically uses the hydrodynamic function...
متن کاملAccurate Calibration and Uncertainty Estimation of the Normal Spring Constant of Various AFM Cantilevers
Measurement of force on a micro- or nano-Newton scale is important when exploring the mechanical properties of materials in the biophysics and nanomechanical fields. The atomic force microscope (AFM) is widely used in microforce measurement. The cantilever probe works as an AFM force sensor, and the spring constant of the cantilever is of great significance to the accuracy of the measurement re...
متن کاملCalibration of rectangular atomic force microscope cantilevers
A method to determine the spring constant of a rectangular atomic force microscope cantilever is proposed that relies solely on the measurement of the resonant frequency and quality factor of the cantilever in fluid ~typically air!, and knowledge of its plan view dimensions. This method gives very good accuracy and improves upon the previous formulation by Sader et al. @Rev. Sci. Instrum. 66, 3...