Shaving of a Frova Intubating Catheter

نویسنده

  • Fergus Magill
چکیده

We report a complication that occurred when using a frova intubating introducer to assist placement of a Mallinckrodt double lumen tube (DLT) in a patient undergoing oesphagectomy. The non compatibility of this commonly used airway device with the DLT resulted in plastic shavings from the introducer contaminating the patients bronchial tree. All fragments of plastic were retrieved and the patient suffered no harm as a result.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Combined use of a McGrath® MAC video laryngoscope and Frova Intubating Introducer in a patient with Pierre Robin syndrome: a case report

Patients with Pierre Robin syndrome are characterized by micrognathia, retrognathia, glossoptosis, and respiratory obstruction and are prone to have a difficult-to-intubate airway. The McGrath® MAC video laryngoscope provides a better view of the glottis than a Macintosh laryngoscope, but it is not easy to insert an endotracheal tube through the vocal cords because a video laryngoscope has a mu...

متن کامل

Massive haemoptysis from a bougie intubating catheter in a patient with endobronchial sarcoid.

To cite: Alter M, Peake B, Grodski S, et al. BMJ Case Rep Published online: [please include Day Month Year] doi:10.1136/bcr-2016215936 DESCRIPTION A 61-year-old man with biopsy-proven sarcoidosis on prednisolone (10 mg daily) presented for elective hemithyroidectomy for a retrosternal goitre. Preoperative investigations revealed bilateral hilar adenopathy on chest radiograph, with preoperative ...

متن کامل

Novel use of an exchange catheter to facilitate intubation with an Aintree catheter in a tall patient with a predicted difficult airway: a case report

INTRODUCTION The Aintree intubating catheter (Cook® Medical Inc., Bloomington, IN, USA) has been shown to successfully facilitate difficult intubations when other methods have failed. The Aintree intubating catheter (Cook® Medical Inc., Bloomington, IN, USA) has a fixed length of 56 cm, and it has been suggested in the literature that it may be too short for safe use in patients who are tall. ...

متن کامل

Retrograde intubation using a frova intubating introducer in a patient with a tracheostomy.

1. Hebl J, Lennon R. Mayo Clinic Atlas of Regional Anesthesia and Ultrasound-Guided Nerve Blockade. 1st ed. 2010: New York: Oxford University Press, 2010 2. Mulroy MF. Book review: Mayo Clinic Atlas of Regional Anesthesia and Ultrasound-Guided Nerve Block. Anesth Analg 2011;113:962 3. Ludwin DB. Book review: Mayo Clinic Atlas of Regional Anesthesia and Ultrasound-Guided Nerve Blockade. Anesth A...

متن کامل

Cuff extrusion in peritoneal dialysis: single-centre experience with the cuff-shaving procedure in five patients over a 4-year period

Catheter-related infections in peritoneal dialysis (PD) remain a significant complication, and some patients with recurrent exit-site (ESI) and/or tunnel infections may experience external cuff extrusion. In these cases, cuff-shaving has been described as a possible course of treatment. During a 4-year period, there were 44 patients with PD at our department; all received double-cuffed Tenckhof...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2015