Another physician's office analyzer: the Abbott "Vision" evaluated.
نویسندگان
چکیده
We evaluated the Abbott Vision system, to assess its suitability for use in the physician's office setting. We compared results from Vision for all the available analytes (alkaline phosphatase, urea nitrogen, cholesterol, creatinine, glucose, uric acid, and triglycerides) with results from the Kodak Ektachem 400 and 700. For all analytes, standard curve linearity, assessed with various dilutions of an analyte-supplemented serum pool, was within the ranges claimed by the manufacturer. Within-run precision (CV) for assays of these analytes ranged from 0.9% to 4.7%, run-to-run precision from 1.1% to 7.3%. Comparisons with other methods were generally very good except for a bias in results for blood urea nitrogen and alkaline phosphatase. Hemolysis (hemoglobin at 2.0 and 3.0 g/L) interferes with results for cholesterol, glucose, triglycerides, and uric acid in serum and whole blood. Bilirubin at 82 mg/L interferes with results for creatinine and triglycerides; at 120 mg/L it interferes with cholesterol, glucose, and uric acid results; and at 170 mg/L it interferes with alkaline phosphatase results. Triglycerides up to 5000 mg/L do not interfere with any of the tests. Calibration of the analyzer was stable for one month. We also compared the performance of a skilled operator with that of an unskilled operator and a physician.
منابع مشابه
Accuracy and precision of methods for theophylline measurement in physicians' offices.
We tested the accuracy and precision of five theophylline methods intended for use in physicians' offices. The Syntex AccuLevel, Ames Seralyzer, and 3M Diagnostics TheoFAST methods were less reproducible (CVs 6.3% to 9.2%) than the Abbott Vision and Kodak DT-60 (CVs 2.2% to 3.3%). Caffeine interfered with the Vision, Seralyzer, and AccuLevel methods, and theobromine interfered with the Vision, ...
متن کاملA physician's office-based digoxin assay (Seralyzer) evaluated for interference by endogenous digoxin-like immunoreactive factors.
A digoxin test for a physician's office based-chemistry analyzer (Ames Seralyzer) was evaluated for possible interference by digoxin-like immunoreactive factors (DLIF). Sera from patients likely to have high concentrations of DLIF (renal and hepatic patients, pregnant women, and neonates) as well as from normal patients and umbilical cord blood were analysed by the Seralyzer digoxin immunoassay...
متن کاملMeasurement of six enzymes with the Kodak DTSC Module, a physician's office analyzer.
We evaluated the performance of the Kodak DTSC Module for determination of alanine aminotransferase (ALT; EC 2.6.1.2), aspartate aminotransferase (2.5.1.2), alkaline phosphatase (3.1.3.1), creatine kinase (2.7.3.2), gamma-glutamyltransferase (2.3.2.2), and lactate dehydrogenase (1.1.1.27). The DTSC is a "special chemistry" accessory for the DT60 analyzer; the same multilayer film technology as ...
متن کاملThe theophylline method of the Abbott "Vision" analyzer evaluated.
We evaluated the analytical performance of the Abbott "Vision" analyzer for theophylline measurement. The within-day precision (CV) was 1.8% and 3.1% at theophylline concentrations of 15.2 and 25.2 mg/L, respectively; between-day precision was 3.5% and 4.8% at 14.9 and 24.4 mg/L, respectively. Bilirubin (143 mg/L) and triglyceride (7.4 g/L) did not interfere, but hemoglobin caused lower values ...
متن کاملBias and precision of cholesterol analysis by physician's office analyzers.
We studied the bias and precision of serum cholesterol analysis by physician's office analyzers. Total imprecision (CV range, %) for analysis of serum pools was: Abbott Vision 1.5%-1.9%; Ames Seralyzer 3.9%-4.5%; BMD Reflotron 2.3%-3.8%; Chrometrics Cholesterol Test System 2.3%-2.8%; Kodak DT-60 1.6%-2.2%. The Ames Seralyzer exhibited an excessive between-run component of variation. We collecte...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 33 6 شماره
صفحات -
تاریخ انتشار 1987