Percutaneous Retrieval of an Embolized Kyphoplasty Cement Fragment From the Pulmonary Artery: A Case Report and Literature Review.
نویسندگان
چکیده
We present a case of a 41-year-old man with symptomatic pulmonary cement embolism following percutaneous vertebral augmentation, which was successfully retrieved via a percutaneous endovascular approach, a novel technique with only two prior cases reported. Cement leakage, including venous embolization of cement into the cardiopulmonary circulation, is a known potential complication following percutaneous kyphoplasty and vertebroplasty. While many patients with pulmonary cement embolism are asymptomatic and likely go undiagnosed, others experience respiratory distress and hemodynamic compromise requiring surgical and medical intervention. The optimal management for pulmonary cement embolism must be tailored to fit each individual patient, dependent upon the acuity of the clinical presentation, coexisting patient comorbidities, and the risks of systemic anticoagulation. In our patient, cement migration was visualized in real-time during vertebral augmentation. Endovascular retrieval by our Interventional Radiology section obviated the need for anticoagulation therapy or more invasive open surgical procedures.
منابع مشابه
Wireless Connection between Guide Wires and Bone Cement: Extravasated Methyl Methacrylate Mimicking a Retained Guide Wire
We present the case of a 56-year-old double lung transplant recipient with chest pain who underwent an attempted endovascular retrieval of what was described as a retained guide wire in the azygos vein. After successfully grasping the tip, the object further migrated to the right pulmonary artery complicating the retrieval. It was realized that the "wire" was extravasated methyl methacrylate fr...
متن کاملPercutaneous Forceps Retrieval of an Embolized Amplatzer Duct Occluder
Transcatheter closure of patent ductus arteriosus with the Amplatzer duct occluder shows excellent results. However, there is a risk of device embolization after deployment. We report the case of a 33-year-old man in whom an embolized Amplatzer device was retrieved from the right pulmonary artery by a percutaneous method. We also review this uncommon severe complication and offer technical tips...
متن کاملUrgent Surgical Intervention for Embolized Cardiac Occluder Devices: A Case Series
Introduction: In this study, we sought to illustrate our experience in urgent surgical management for embolized cardiac septal occlude devices resulting from trans-catheter closure of atrial septal defect and ventricular septal defect. Mathrials and Methods: We retrospectively reviewed four patients aged 2–10 years who underwent urgent surgery due to cardiac septal occluder embolization between...
متن کاملHUGE ANEURYSM OF T H E DUCTUS ARTERIOSUS IN CHILDHOOD: CASE REPORT AND LITERATURE REVIEW
Patent ductus arteriosus (PDA) is an open communication between the upper descending thoracic aorta and the proximal left pulmonary artery, which results from persistence of the fetal ductus arteriosus. It may be isolated or occur with other congenital cardiac anomalies. Aneurysms of the ductus arteriosus are rare lesions which may occur as a spontaneous. infantile fonn or occur in later c...
متن کاملSuccessful Image-Guided Retrieval of an Embolized Fragment of a Fractured Haemodialysis Catheter Tip from the Pulmonary Artery
BACKGROUND Cuffed, double-lumen, tunneled haemodialysis catheters are a common means of vascular access in paediatric haemodialysis, particularly in infants. Haemodialysis catheter fracture with distal embolization is a rare complication. CASE REPORT A 2-year-old boy was receiving chronic haemodialysis via a right internal jugular cuffed, double-lumen, tunneled haemodialysis catheter, inserte...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Journal of radiology case reports
دوره 10 7 شماره
صفحات -
تاریخ انتشار 2016