Increase in Halothane Concentration following Refilling of Certain Vaporizers.
نویسندگان
چکیده
A case of unexpected cardiovascular collapse during nitrous oxide, oxygen and halothane anaesthesia recently occurred in the hospital. A flow rate of 8 l./min was in use via a Magill attachment. Effective treatment was instituted and the patient made a satisfactory recovery. Some days later discussion of the event brought to light the fact that the halothane vaporizer, a calibrated Boyle bottle, had been refilled shortly before the episode. This led to the possibility being considered that an unduly high concentration of halothane might be delivered by a recently refilled vaporizer because of the inevitable splashing of the sides of the bottle with liquid halothane. This possibility had not been thought of at the time, and the interval between refilling and the collapse, therefore, was not noted.
منابع مشابه
A clinical and laboratory evaluation of four fluothane vaporizers.
WIDESPREAD CONTINUING INTEREST in 2 brom, 2 chlor 1, 1, 1, fluoroethane, known as halothane or more commonly Fluothane, has resulted over the past few months in many attempts to make its use sas and at the same time more economical. Certain apparatus has been designed to facilitate its use, particularly by the less experienced anaesthetist. This paper describes the clinical and laboratory inves...
متن کاملThe action of the halothane-diethyl ether azeotropic mixture on experimental animals.
MANY mixtures of volatile anaesthetics have been -used in the past with some success, but they are difficult to handle in the modern anaesthetic apparatus. Because of the differences in the boiling points of the components of these mixtures, they volatilize at different rates and therefore it is almost impossible to know their concentration in the anaesthetic mixture administered to the patient...
متن کاملFlow reversal through a Mark III halothane vaporizer.
A fault in the assembly of a Matrix Large Animal Circle anaesthetic machine resulted in reversal of fresh gas flow through the vaporizer. The fault was discovered only after the sudden development of excessive depth of anaesthesia in two equine patients. Laboratory investigations were conducted to determine the effect of flow reversal on vaporizer output. Results indicated that output concentra...
متن کاملAnesthesia-related cardiac arrest in children.
To the Editor:—Morray et al. are to be congratulated in trying to sort out some of the issues regarding intraoperative cardiac arrest in children. However, I am concerned with the conclusions regarding cardiac arrests related to anesthetic agents. The authors freely admit that they do not know the numerator or denominator and therefore cannot say that sevoflurane is safer than halothane, but th...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- British journal of anaesthesia
دوره 37 شماره
صفحات -
تاریخ انتشار 1965