On-line monitoring of intrinsic PEEP in ventilator-dependent patients.

نویسندگان

  • G Nucci
  • M Mergoni
  • C Bricchi
  • G Polese
  • C Cobelli
  • A Rossi
چکیده

Measurement of the intrinsic positive end-expiratory pressure (PEEP(i)) is important in planning the management of ventilated patients. Here, a new recursive least squares method for on-line monitoring of PEEP(i) is proposed for mechanically ventilated patients. The procedure is based on the first-order model of respiratory mechanics applied to experimental measurements obtained from eight ventilator-dependent patients ventilated with four different ventilatory modes. The model PEEP(i) (PEEP(i,mod)) was recursively constructed on an inspiration-by-inspiration basis. The results were compared with two well-established techniques to assess PEEP(i): end-expiratory occlusion to measure static PEEP(i) (PEEP(i, st)) and change in airway pressure preceding the onset of inspiratory airflow to measure dynamic PEEP(i) (PEEP(i,dyn)). PEEP(i, mod) was significantly correlated with both PEEP(i,dyn) (r = 0.77) and PEEP(i,st) (r = 0.90). PEEP(i,mod) (5.6 +/- 3.4 cmH(2)O) was systematically >PEEP(i,dyn) and PEEP(i,st) (2.7 +/- 1.9 and 8.1 +/- 5.5 cmH(2)O, respectively), in all the models without external PEEP. Focusing on the five patients with chronic obstructive pulmonary disease, PEEP(i,mod) was significantly correlated with PEEP(i,st) (r = 0.71), whereas PEEP(i,dyn) (r = 0.22) was not. When PEEP was set 5 cmH(2)O above PEEP(i,st), all the methods correctly estimated total PEEP, i.e., 11.8 +/- 5.3, 12.5 +/- 5.0, and 12.0 +/- 4.7 cmH(2)O for PEEP(i,mod), PEEP(i,st), and PEEP(i,dyn), respectively, and were highly correlated (0.97-0.99). We interpreted PEEP(i,mod) as the lower bound of PEEP(i,st) and concluded that our method is suitable for on-line monitoring of PEEP(i) in mechanically ventilated patients.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Measurement of air trapping, intrinsic positive end-expiratory pressure, and dynamic hyperinflation in mechanically ventilated patients.

Severe airflow obstruction is a common cause of acute respiratory failure. Dynamic hyperinflation affects tidal ventilation, increases airways resistance, and causes intrinsic positive end-expiratory pressure (auto-PEEP). Most patients with asthma and chronic obstructive pulmonary disease have dynamic hyperinflation and auto-PEEP during mechanical ventilation, which can cause hemodynamic compro...

متن کامل

Effects of PEEP on respiratory mechanics in patients with COPD on mechanical ventilation.

We studied the effects of positive end-expiratory pressure (PEEP) applied by the ventilator on respiratory mechanics in ventilated patients with chronic obstructive pulmonary disease (COPD). Airway pressures, relaxed expiratory flow-volume curves and end-expiratory volumes (EEV) were measured. In all patients investigated without PEEP applied by the ventilator, an intrinsic PEEP level (PEEPi) a...

متن کامل

Bedside measurement of changes in lung impedance to monitor alveolar ventilation in dependent and non-dependent parts by electrical impedance tomography during a positive end-expiratory pressure trial in mechanically ventilated intensive care unit patients

INTRODUCTION As it becomes clear that mechanical ventilation can exaggerate lung injury, individual titration of ventilator settings is of special interest. Electrical impedance tomography (EIT) has been proposed as a bedside, regional monitoring tool to guide these settings. In the present study we evaluate the use of ventilation distribution change maps (DeltafEIT maps) in intensive care unit...

متن کامل

Positive end-expiratory pressure during mechanical ventilation and respiratory support in newborns and children

Positive end-expiratory pressure (PEEP) is used during non-invasive and invasive ventilation of newborns, infants and children. PEEP improves gas exchange by increasing the functional residual capacity, reducing respiratory effort, lowering requirements for respiratory mixture oxygen, and enabling a decrease in the peak inspiratory pressure (PIP) without decreasing the mean airway pressure. Its...

متن کامل

PROCEEDINGS 90th ANNUAL MEETING 43 DESIGN OF AUTOMATED DETECTION OF INCOMPLETE EMPTYING FROM VENTILATOR GRAPHICS: EVALUATING THRESHOLD

DESIGN OF AUTOMATED DETECTION OF INCOMPLETE EMPTYING FROM VENTILATOR GRAPHICS: EVALUATING THRESHOLD VALUES. Nyimas Y. Isti Arief, Curtis N. Sessler, Paul A. Wetzel, and Mary Jo E. Grap, Dept. of Biomedical Engineering, Virginia Commonwealth University. Auto-PEEP stemming from incomplete emptying of breath inhalation is an undesired excessive pressure in the lungs of patients on invasive mechani...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • Journal of applied physiology

دوره 89 3  شماره 

صفحات  -

تاریخ انتشار 2000