Cardiac death and stored electrograms in patients with third-generation implantable cardioverter-defibrillators.
نویسندگان
چکیده
OBJECTIVES We sought to utilize terminal stored intracardiac electrograms (EGMs) to study the electrophysiologic events that accompany mortality in patients with third-generation implantable cardioverter-defibrillators (ICDs). BACKGROUND Despite their ability to effectively terminate ventricular tachyarrhythmias, cardiac mortality in patients with ICDs remains high. The mechanisms and modes of death in these patients are not well understood. METHODS We retrospectively analyzed clinical data and stored EGMs from patients enrolled in the clinical trial of the Ventritex Cadence ICD. Of the 1,729 patients 119 died during 6 years of follow-up. The final recorded EGM was reviewed. Postimplant EGMs as well as 50 control EGMs were used to define normal EGM characteristics. RESULTS There were 36 noncardiac deaths (30%) and 83 cardiac deaths (70%). Of the cardiac deaths, 55 (66%) were nonsudden and 28 (34%) were sudden. When cardiac deaths were analyzed, 46 (55%) had no stored EGMs within 1 h of death, implying that the deaths were not directly related to tachyarrhythmias. In 37 cardiac deaths (18 nonsudden, 19 sudden), stored EGMs were present within 1 h of death. In these 37 deaths, the final EGM recorded was wide (>158 ms) in 33 (89%). Wide EGMs were interpreted as ventricular tachycardia in 27 and ventricular fibrillation in 6. In 13 of the 33 patients (39%) with wide EGMs, therapy was not delivered by the ICD, as it incorrectly detected a spontaneous termination of the arrhythmia. EGMs were significantly wider if recorded within 1 h, as compared with those recorded from 1 to 48 h before death (261+/-124 vs. 181+/-93 ms, p=0.04). CONCLUSIONS Only 37 patients (31%) who died after placement of an ICD had a stored EGM within 1 h of the time of death, suggesting that the majority of deaths (69%) were not the immediate result of a tachyarrhythmia. When EGMs were recorded, they were wide in 89% of patients. These wide EGMs most likely represent intracardiac recordings of electromechanical dissociation. Thus, of the 119 deaths, 112 (94%) were not the immediate result of a tachyarrhythmia.
منابع مشابه
Failure of an implantable cardioverter-defibrillator to redetect ventricular fibrillation in patients with a nonthoracotomy lead system.
BACKGROUND Shock delivery of an implantable defibrillator may cause a change in the amplitude of endocardial electrograms and impair the detection of ventricular fibrillation. Thus, the effects of shock discharges on the amplitude of endocardial electrograms were evaluated in five patients undergoing implantation of a cardioverter-defibrillator in combination with a new nonthoracotomy lead syst...
متن کاملLead System
Background. Shock delivery of an implantable defibrillator may cause a change in the amplitude of endocardial electrograms and impair the detection of ventricular fibrillation. Thus, the effects of shock discharges on the amplitude of endocardial electrograms were evaluated in five patients undergoing implantation of a cardioverter-defibrillator in combination with a new nonthoracotomy lead sys...
متن کاملDeceptive intracardiac electrograms during an ICD interrogation. A clinical challenge
We present a case of a patient with significant structural heart disease from previous cardiac surgeries who presented for an electrophysiology study and radiofrequency ablation. The case highlights the significance of intra-atrial conduction delays in such patients and potential pitfalls it can present while interpreting intra-cardiac electrograms especially during implantable cardioverter def...
متن کاملAn automated, web-enabled and searchable database system for archiving electrogram and related data from implantable cardioverter defibrillators.
The advent of implantable cardioverter defibrillators (ICDs) has resulted in significant reductions in mortality in patients at high risk for sudden cardiac death. Extensive related basic research and clinical investigation continue. ICDs typically record intracardiac electrograms and inter-beat intervals along with device settings during episodes of device delivery of therapy. Researchers wish...
متن کاملLife expectancy gains and cost-effectiveness of implantable cardioverter/defibrillators for the primary prevention of sudden cardiac death in patients with hypertrophic cardiomyopathy
CRD summary The study examined the cost-effectiveness of using implantable cardioverter/defibrillator (ICD) insertion or amiodarone therapy for the prevention of sudden cardiac death (SCD) in patients with hypertrophic cardiomyopathy (HCM), with no history of cardiac arrest but a significant risk of SCD. The study results support the cost-effectiveness of ICD insertion from the perspective of t...