A comparative evaluation of the analytical performances of Capillarys 2 Flex Piercing, Tosoh HLC-723 G8, Premier Hb9210, and Roche Cobas c501 Tina-quant Gen 2 analyzers for HbA1c determination
نویسندگان
چکیده
INTRODUCTION Haemoglobin A1c (HbA1c) is widely used in the management of diabetes. Therefore, the reliability and comparability among different analytical methods for its detection have become very important. MATERIALS AND METHODS A comparative evaluation of the analytical performances (precision, linearity, accuracy, method comparison, and interferences including bilirubin, triglyceride, cholesterol, labile HbA1c (LA1c), vitamin C, aspirin, fetal haemoglobin (HbF), and haemoglobin E (Hb E)) were performed on Capillarys 2 Flex Piercing (Capillarys 2FP) (Sebia, France), Tosoh HLC-723 G8 (Tosoh G8) (Tosoh, Japan), Premier Hb9210 (Trinity Biotech, Ireland) and Roche Cobas c501 (Roche c501) (Roche Diagnostics, Germany). RESULTS A good precision was shown at both low and high HbA1c levels on all four systems, with all individual CVs below 2% (IFCC units) or 1.5% (NGSP units). Linearity analysis for each analyzer had achieved a good correlation coefficient (R2 > 0.99) over the entire range tested. The analytical bias of the four systems against the IFCC targets was less than ± 6% (NGSP units), indicating a good accuracy. Method comparison showed a great correlation and agreement between methods. Very high levels of triglycerides and cholesterol (≥ 15.28 and ≥ 8.72 mmol/L, respectively) led to falsely low HbA1c concentrations on Roche c501. Elevated HbF induced false HbA1c detection on Capillarys 2FP (> 10%), Tosoh G8 (> 30%), Premier Hb9210 (> 15%), and Roche c501 (> 5%). On Tosoh G8, HbE induced an extra peak on chromatogram, and significantly lower results were reported. CONCLUSIONS The four HbA1c methods commonly used with commercial analyzers showed a good reliability and comparability, although some interference may falsely alter the result.
منابع مشابه
Analytical verification and quality assessment of the Tosoh HLC-723GX HbA1c analyzer
OBJECTIVES Ion-exchange high-performance liquid chromatography (IE-HPLC) has long been used as a reproducible and versatile analytical tool for HbA1c measurement. In this study, we performed analytical verification and quality assessment of the recently introduced small IE-HPLC Tosoh HLC-723GX HbA1c analyzer, and a comparison of results to immunoassay (IA) and capillary electrophoresis (CE). ...
متن کاملMeasurement of HbA1c and HbA2 by Capillarys 2 Flex Piercing HbA1c programme for simultaneous management of diabetes and screening for thalassemia
INTRODUCTION Thalassemia could interfere with some assays for haemoglobin A1c (HbA1c) measurement, therefore, it is useful to be able to screen for thalassemia while measuring HbA1c. We used Capillarys 2 Flex Piercing (Capillarys 2FP) HbA1c programme to simultaneously measure HbA1c and screen for thalassemia. MATERIALS AND METHODS Samples from 498 normal controls and 175 thalassemia patients ...
متن کاملAnalytical performance of 4 automated assays for measurement of cystatin C.
BACKGROUND We evaluated the analytical performance of 4 cystatin C assays (Siemens N Latex on BNII, Roche Tina-quant on Cobas c501, Genzyme on Cobas c501, and Tosoh ST AIA-PACK on Tosoh AIA-600II) according to guidelines published by the Clinical and Laboratory Standards Institute. METHODS We evaluated total imprecision, limit of detection, and limit of quantification for each assay using pat...
متن کاملMulticenter evaluation of Tosoh glycohemoglobin analyzer.
BACKGROUND We describe an Anglo-French evaluation of a new analyzer. METHODS The Tosoh HLC-723 GHb V, A1c2.2 glycohemoglobin analyzer is an HPLC instrument with primary blood tube sampling, bar-code reading, cap piercing, and the ability to chromatographically separate labile hemoglobin A1c (HbA1c). We evaluated two analytical protocols, 2.2 and 3.0 min, and compared results for blood samples...
متن کامل