Optical Filter Transmittance Measurements by a Double-Beam Spectrophotometer for Calibration of Fluorescence Detector AUGER Filter Samples

نویسندگان

  • A. Geranios
  • E. Fokitis
  • A. Petridis
  • M. Vassiliou
چکیده

An experimental setup and procedure for accurate transmittance measurements of candidate filters for the Auger Fluorescence Detector is presented. For this check, we used the Hitachi U-2000 double-beam spectrophotometer (Sph) of the Chemical Department of the University of Athens (UoA). After the calibration of the Sph, we are able to observe transmittance deviations over the area of the filter under study within a sensitivity better than 0.2%. A set of candidate filters is measured to identify their spectral transmittance in response to different angles of incidence. Also, the use of the Sph for few layers developmental multilayer can be studied for their spectral transmittance.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Optical Filter Design for the Fluorescence Detector of the AUGER Project

The design concept of the required optical filters for the Fluorescence Detector of the AUGER Project, is presented in this work. Due to the needed high transmittance values in the UV region, the dielectric thin film multilayer technique, using UV-transparent materials, has been selected. The design method applied, is the “Monte Carlo Simulated Annealing” (MOCASA). Two alternative target spectr...

متن کامل

An Experimental Method to Investigate the Optimal UV Filter for the Auger Fluorescence Detectors

A specific experimental method, in order to investigate the more suitable UV filter to install in the Fluorescence Detector (FD) of the AUGER Project, is presented in this work. A prototype experimental setup has been developed for this purpose. As detector, it uses an elementary cell, similar to the one used in the FD. A fast and sensitive photon counting system, is used for the light intensit...

متن کامل

Multi-wavelength Calibration Procedure for the Pierre Auger Observatory Fluorescence Detectors

We present a method to measure the relative spectral response of the Pierre Auger Observatory Fluorescence Detector. The calibration was done at wavelengths of 320, 337, 355, 380 and 405 nm using an end-to-end technique in which the response of all detector components are combined in a single measurement. A xenon flasher and notch–filters were used as the light source for the calibration device...

متن کامل

Multi-wavelength Calibration Procedure for the Pierre Auger Observatory Fluorescence Detectors

We present a method to measure the relative spectral response of the Pierre Auger Observatory Fluorescence Detector. The calibration was done at wavelengths of 320, 337, 355, 380 and 405 nm using an end-to-end technique in which the response of all detector components are combined in a single measurement. A xenon flasher and notch–filters were used as the light source for the calibration device...

متن کامل

A fluorescence detector for high-pressure liquid chromatography.

We describe a fluorescence spectrophotometer adapted with a micro quartz flow cell to record the output of modern liquid chromatographs. The optical system is double beam in that the light source variations are cancelled out by a second photomultiplier, thus enhancing the sensitivity of the technique. The emission spectra may be scanned by stopping the flow in the chromatographic column and sca...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره   شماره 

صفحات  -

تاریخ انتشار 2000