A Comprehensive Assessment of the Precision and Agreement of Anterior Corneal Power Measurements Obtained Using 8 Different Devices

نویسندگان

  • Qinmei Wang
  • Giacomo Savini
  • Kenneth J. Hoffer
  • Zhen Xu
  • Yifan Feng
  • Daizong Wen
  • Yanjun Hua
  • Feng Yang
  • Chao Pan
  • Jinhai Huang
چکیده

PURPOSE To comprehensively assess the precision and agreement of anterior corneal power measurements using 8 different devices. METHODS Thirty-five eyes from 35 healthy subjects were included in the prospective study. In the first session, a single examiner performed on each subject randomly measurements with the RC-5000 (Tomey Corp., Japan), KR-8000 (Topcon, Japan), IOLMaster (Carl Zeiss Meditec, Germany), E300 (Medmont International, Australia), Allegro Topolyzer (Wavelight AG, Germany), Vista (EyeSys, TX), Pentacam (Oculus, Germany) and Sirius (CSO, Italy). Measurements were repeated in the second session (1 to 2 weeks later). Repeatability and reproducibility of corneal power measurements were assessed based on the intrasession and intersession within-subject standard deviation (Sw), repeatability (2.77Sw), coefficient of variation (COV), and intraclass correlation coefficient (ICC). Agreement was evaluated by 95% limits of agreement (LoA). RESULTS All devices demonstrated high repeatability and reproducibility of the keratometric values (2.77Sw<0.36D, COV<0.3%, ICC>0.98). Repeated-measures analysis of variance with Bonferroni post test showed statistically significant differences (P<0.01) among mean keratometric values of most instruments; the largest differences were observed between the EyeSys Vista and Medmont E300. Good agreement (i.e., 95%LoA within ± 0.5D) was found between most instruments for flat, steep and mean keratometry, except for EyeSys and Medmont. Repeatability and reproducibility of vectors J(0) and J(45) was good, as the ICCs were higher than 0.9, except J(45) of Medmont and Pentacam. For the 95% LoAs of J(0) and J(45), they were all ≤ ± 0.31 among any two paired devices. CONCLUSIONS The 8 devices showed excellent repeatability and reproducibility. The results obtained using the RC-5000, KR-8000, IOLMaster, Allegro Topolyzer, Pentacam and Sirius were comparable, suggesting that they could be used interchangeably in most clinical settings. Caution is warranted with the measurements of the EyeSys Vista and Medmont E300, which should not be used interchangeably with other devices due to lower agreement. TRIAL REGISTRATION ClinicalTrials.gov NCT01587287.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

مقایسه‌ی کراتومتری پنتاکم وکراتومتری IOL Master در محاسبه‌ی لنز داخل چشمی افراد با قرنیه‌ی طبیعی

Background and Objective: Proper method and machine for corneal evaluation is an important factor in many anterior segment interventions. This study was performed to compare the mean keratometry (K) readings obtained with an IOL Master and Pentacam for intraocular lens power calculation in normal subjects with no history of refractive surgery. Materials and Methods: Mean K values were obtained ...

متن کامل

Precision and Agreement of Corneal Power Measurements Obtained Using a New Corneal Topographer OphthaTOP

PURPOSE To evaluate repeatability and reproducibility of anterior corneal power measurements obtained with a new corneal topographer OphthaTOP (Hummel AG, Germany) and agreement with measurements by a rotating Scheimpflug camera (Pentacam HR, Oculus, Germany) and an automated keratometer (IOLMaster, Carl Zeiss Meditec, Germany). METHODS The right eyes of 79 healthy subjects were prospectively...

متن کامل

Comparison between Pentacam HR and Orbscan II after Hyperopic Photorefractive Keratectomy

PURPOSE The aim of this study was to determine the agreement between Pentacam HR (Scheimpflug imaging, Oculus) and Orbscan II (scanning slit topography, Bausch and Lomb) in measuring corneal parameters after photorefractive keratectomy (PRK) for hyperopia. METHODS In this prospective cross-sectional study, 38 hyperopic eyes undergoing PRK were examined before refractive surgery and 8 to 10 mo...

متن کامل

Assessment of anterior and posterior corneal indices using two Scheimpflug analyzers.

PURPOSE To assess the agreement between the elevation and curvature measurements of the anterior and posterior corneal surfaces obtained using the Galilei Dual Scheimpflug Analyzer and those obtained using the Pentacam single Scheimpflug system. METHODS This prospective, noninterventional, diagnostic study was conducted at the Department of Ophthalmology at the American University of Beirut (...

متن کامل

Comparison of Central Corneal Thickness Measured through Pentacam, OCTCasia2, and Orbscan IIz in Normal Individuals

Purpose: To compare the central corneal thickness measured through Pentacam, OCTCasia2, and Orbscan IIz in normal individuals. Methods: In this cross sectional comparative study, the researcher investigated the eyes of 70 normal candidates (140 eyes) with the mean age of 31.42 ± 7.70, referring to Dr. Sedaghat eye clinic for refractive surgery. For each person, Central Corneal Thickness (CCT) w...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:

دوره 7  شماره 

صفحات  -

تاریخ انتشار 2012