Mini-Wright peak flow meters are reliable after 5 years use. Dutch CNSLD Study Group.

نویسندگان

  • W R Douma
  • T W van der Mark
  • H T Folgering
  • E Kort
  • H A Kerstjens
  • G H Koëter
  • D S Postma
چکیده

In studies with a long-term follow-up, peak expiratory flow (PEF) meters are often used to assess bronchial obstruction. The question arises whether data obtained with these frequently used meters are still reliable after several years of use, and whether the old meters should be renewed after a certain period. In the present study, we tested the reliability of PEF values measured with mini-Wright PEF meters that had been used frequently for 5 yrs. The values obtained with these meters were compared with values measured with identical but new meters, in 50 patients with obstructive airways disease. Though statistically significant, there was no clinically significant difference in mean PEF measured with the old and new meters in most patients (mean difference 10.2 L.min-1). However, on an individual basis, the differences between old and new meters could be large (upper and lower limits of agreement (mean +/- 2 SD) being 63.6 and -43.2 L.min-1, respectively). We conclude that mean peak expiratory flow values measured with frequently used mini-Wright peak expiratory flow meters are still reliable after 5 yrs. In long-term studies, renewal of peak expiratory flow meters should be restricted to cases of obvious malfunction.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Bovine TB in badgers: a reappraisal of aetiology and pathogenesis.

I read with great interest the article of Miles et al. (1) about the performance of mini Wright peak flow meters after prolonged use. We performed a study (2) comparing the reliability of five new mini Wright peak flow meters before and after 10 000 compressed air impulses of about 4.50 1 min ’ flow, controlled by an electronically driven magnetic valve. There were no significant deviations of ...

متن کامل

Changes in peak expiratory flow indices as a proxy for changes in bronchial hyperresponsiveness. Dutch Chronic Non-Specific Lung Disease study group.

Guidelines for asthma management advocate home peak expiratory flow (PEF) monitoring. It is commonly stated that PEF variability is a good proxy of bronchial hyperresponsiveness (BHR), a hallmark of asthma. However, this has hardly been tested longitudinally, as required to monitor asthma. This study assesses which PEF index correlates best with BHR longitudinally and whether the correlation im...

متن کامل

Calibration of some Wright peak flow meters.

The performance of Wright peak flow meters is described using an absolute and precise calibration method. This showed errors as great as 25% with the peak flow meter and 40% with the mini peak flow meter. At large flow rates errors were less. The expected variation between individual peak flow meters was about 12%. The expected variation between individual mini peak flow meters was between 16% ...

متن کامل

Measuring peak expiratory flow in general practice: comparison of mini Wright peak flow meter and turbine spirometer.

OBJECTIVE To compare measurements of the peak expiratory flow rate taken by the mini Wright peak flow meter and the turbine spirometer. DESIGN Pragmatic study with randomised order of use of recording instruments. Phase 1 compared a peak expiratory flow type expiration recorded by the mini Wright peak flow meter with an expiration to forced vital capacity recorded by the turbine spirometer. P...

متن کامل

ذخیره در منابع من


  با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

عنوان ژورنال:
  • The European respiratory journal

دوره 10 2  شماره 

صفحات  -

تاریخ انتشار 1997