Analytical and clinical performance characteristics of Tandem-MP Ostase, a new immunoassay for serum bone alkaline phosphatase.
نویسندگان
چکیده
The performance characteristics of the Tandem-MP Ostase assay, a new microplate immunoassay for bone-specific alkaline phosphatase (bone ALP; EC 3.1.3.1) in human sera, are described. Bone ALP is bound to streptavidin-coated microwells by a single biotinylated anti-bone ALP monoclonal antibody. Antigen is detected by the addition of p-nitrophenyl phosphate. The assay is performed at room temperature in <90 min. Imprecision was 2.3-6.1% with a detection limit of 0.6 microg/L. Method comparison of bone ALP measurements with the Tandem-MP Ostase assay and the mass-based Tandem-R Ostase assay (n = 285) indicated regression statistics of Tandem-MP Ostase = 1.03 Tandem-R Ostase + 0.22 microg/L, S(y/x) = 4.0 microg/L, r = 0.97. Serum bone ALP values in apparently healthy men and in pre- and postmenopausal women were also similar between the two Ostase assay formats. Liver ALP reactivity determined using the slope and heat inactivation methods was similar in both Ostase assays. Liver ALP reactivity ranged from 3 microg/L (heat inactivation) to 6 microg/L (slope method) per 100 U/L of liver ALP activity, whereas bone ALP reactivity was 37 microg/L per 100 U/L of bone ALP activity, indicating a liver ALP relative reactivity of 8.1-16.2%. Similar results were obtained with the Alkphase-B bone ALP immunoassay. The Tandem-MP Ostase bone ALP assay demonstrated increased concentrations of serum bone ALP in conditions where bone metabolism is increased and showed a rapid, temporal decrease in serum bone ALP in Paget disease patients on bisphosphonate therapy. In conclusion, the Tandem-MP Ostase assay for serum bone ALP is a rapid, simple, robust nonisotopic alternative to the Tandem-R Ostase immunoradiometric assay that provides an accurate and sensitive assessment of bone turnover.
منابع مشابه
Inter-method variability in bone alkaline phosphatase measurement: clinical impact on the management of dialysis patients.
BACKGROUND Bone-specific alkaline phosphatase (BAP) is now recommended to assess bone turnover in hemodialysis (HD) patients. However, little is known about potential variability between methods available to measure BAP. METHODS We measured BAP in 76 HD patients with six different assays (Beckman-Coulter Ostase IRMA, Beckman-Coulter Ostase Access, IDS iSYS Ostase, IDS Ostase enzyme immunoassa...
متن کاملBone alkaline phosphatase isoenzyme in renal osteodystrophy.
Serum total alkaline phosphatase is the most commonly used biochemical marker of bone disease in renal patients, but alkaline phosphatase originates from different organs and sometimes lacks specificity. Bone isoenzyme measurement is considered superior to total alkaline phosphatase for the assessment of bone metabolism. We have studied the value of bone isoenzyme, determined by a new. IRMA (Ta...
متن کاملImmunoradiometric method and electrophoretic system compared for quantifying bone alkaline phosphatase in serum.
Agarose electrophoresis (Isopal, Beckman) and an immunoradiometric assay (IRMA) involving specific monoclonal antibodies (Ostase, Hybritech), two methods for the quantification of serum bone alkaline phosphatase (ALP, EC 3.1.3.1), a marker of osteoblastic activity, were compared in 293 patients: 79 with end-stage renal failure treated with hemodialysis and 214 with malignant disease. Overall co...
متن کاملPlasma total versus bone alkaline phosphatase as markers of bone turnover in hemodialysis patients.
Plasma total versus bone alkaline phosphatase as markers of bone turnover in hemodialysis patients. Plasma bone-specific alkaline phosphatase (bAP) has been demonstrated to be more reliable than total alkaline phosphatases (tAP) in providing information about bone turnover in patients with metabolic bone diseases. This study surveyed 42 hemodialysis patients who underwent a systematic transilia...
متن کاملAnalytical validation of serum bone alkaline phosphatase (BAP OSTASE) on Liaison.
BACKGROUND The goal of this study was to validate the DiaSorin Liaison BAP OSTASE, a new method for measurement of bone alkaline phosphatase (BAP), and to compare this method with the Beckman-Coulter Access Ostase. We also wanted to establish the reference range for BAP in adults and children. METHODS We determined the precision, functional sensitivity, recovery, linearity and measurement unc...
متن کاملذخیره در منابع من
با ذخیره ی این منبع در منابع من، دسترسی به آن را برای استفاده های بعدی آسان تر کنید
برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید
ثبت ناماگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید
ورودعنوان ژورنال:
- Clinical chemistry
دوره 44 10 شماره
صفحات -
تاریخ انتشار 1998