Application of Regent mechanical valve in patients with small aortic annulus: 3-year follow-up

نویسندگان

  • Dong Zhao
  • Chunsheng Wang
  • Tao Hong
  • Cuizhen Pan
  • Changfa Guo
چکیده

BACKGROUND Aortic valve replacement (AVR) with a small aortic annulus is always challenging for the cardiac surgeon. In this study, we sought to evaluate the midterm performance of implantation with a 17-mm or 19-mm St. Jude Medical Regent (SJM Regent) mechanical valve in retrospective consecutive cohort of patients with small aortic annulus (diameter ≤ 19 mm). METHODS From January 2008 to April 2011, 40 patients (31 female, mean age = 47.2 ± 5.8 years) with small aortic annulus (≤19 mm in diameter) underwent aortic valve replacement with a 17-mm or 19-mm St. Jude Medical Regent (SJM Regent) mechanical valve. Preoperative mean body surface area, New York Heart Association class, and mean aortic annulus were 1.61 ± 0.26 m2, 3.2 ± 0.4, and 18 ± 1.4 mm respectively. Patients were divided into two groups, according to the implantation of 17 mm SJM Regent mechanical valve (group 1, n = 18) or 19 mm SJM Regent valve (group 2, n = 22). All patients underwent echocardiography examination preoperatively and at one year post-operation. RESULTS There were no early deaths in either group. Follow-up time averaged 36 ± 17.6 months. The mean postoperative New York Heart Association class was 1.3 ± 0.6 (p < 0.001). By echocardiography, in group 1, the left ventricular ejection fraction (LVEF), left ventricular fraction shortening (LVFS), and the indexed effective orifice area (EOAI) increased from 43.7% ± 11.6%, 27.3% ± 7.6%, and 0.70 ± 0.06 cm2/m2 to 69.8 ± 9.3%, 41.4 ± 8.3%, and 0.92 ± 0.10 cm2/m2 respectively (P < 0.05), while the left ventricular mass index (LVMI), and the aortic transvalvular pressure gradient decreased from 116.4 ± 25.4 g/m2, 46.1 ± 8.5 mmHg to 86.7 ± 18.2 g/m2, 13.7 ± 5.2 mmHg respectively. In group 2, the LVEF, LVFS and EOAI increased from 45.9% ± 9.7%, 30.7% ± 8.0%, and 0.81 ± 0.09 cm2/m2 to 77.4% ± 9.7%, 44.5% ± 9.6%, and 1.27 ± 0.11 cm2/m2 respectively, while the LVMI, and the aortic transvalvular pressure gradient decreased from 118.3 ± 27.6 g/m2, 44.0 ± 6.7 mmHg to 80.1 ± 19.7 g/m2, 10.8 ± 4.1 mmHg as well. The prevalence of PPM was documented in 2 patients in Group 1. CONCLUSIONS Patients with small aortic annulus and body surface area, experienced satisfactory clinical improvement after aortic valve replacement with modern SJM Regent bileaflet prostheses.

برای دانلود متن کامل این مقاله و بیش از 32 میلیون مقاله دیگر ابتدا ثبت نام کنید

ثبت نام

اگر عضو سایت هستید لطفا وارد حساب کاربری خود شوید

منابع مشابه

Seventeen-millimeter St. Jude Medical Regent valve in patients with small aortic annulus: dose moderate prosthesis-patient mismatch matter?

BACKGROUND The study was designed to evaluate the effects of moderate prosthesis-patient mismatch (defined as 0.65 cm(2)/m(2) <indexed effective orifice area ≤ 0.85 cm(2)/m(2)) on midterm outcomes after isolated aortic valve replacement with a 17-mm St. Jude Medical Regent valve in a large series of patients, and to determine if these effects are influenced by patient confounding variables. M...

متن کامل

A case of aortic valve replacement with St. Jude Medical Regent valve (first implant in Japan).

The St. Jude Medical (SJM) Regent valve was developed as a new mechanical valve by improving the conventional SJM valve. The effective orifice area is wider than that of Hemodynamic Plus (HP) series. The efficacy of the new valve has been reported in Europe and the United States. On October 26, 2004, we first performed aortic valve replacement (AVR) with the SJM Regent valve for aortic valve st...

متن کامل

Early and mid-term outcome in terms of functional and hemodynamic performance of the st. Jude regent 19-mm aortic mechanical prosthesis versus 19-mm carpentier edwards aortic biological prosthesis

BACKGROUND The aim of the present study is to compare the early and mid-term clinical and hemodynamic results of the aortic valve replacement (AVR) with a St Jude Medical Regent 19-mm prosthesis (SJMR-19) versus Carpentied-Edwars bovine pericardial 19-mm valve (CE-19). METHODS Between January 2002 and January 2012, 265 patients (Group I) and 58 patients (Group II) with underwent AVR with a SJ...

متن کامل

AN INNOVATIVE SURGIC AL TECHNIQUE FOR CONTINUOUS SUTURING OF PROSTHESIS IN MITR AL VALVE REPL A CEMENT

Mitral valve replacement (MVR) is usually performed in two methods of running and interrupted suturing. In running method, the suturing of the valve is usually with some traction on the annulus and also there is a risk of aortic valve injury. This study was conducted to evaluate a new technique for suturing to avoid these complications and decrease the time of cardiac arrest. One hundred an...